Measurement of cardioversion/defibrillation thresholds in man by a truncated exponential waveform and an apical patch-superior vena caval spring electrode configuration.

نویسندگان

  • R A Winkle
  • E B Stinson
  • S M Bach
  • D S Echt
  • P Oyer
  • K Armstrong
چکیده

Defibrillation/cardioversion thresholds were measured in 33 patients undergoing defibrillator implants. Each patient had a 12 cm2 patch placed near the left ventricular apex via a thoracotomy and a 10 cm2 spring lead placed pervenously at the right atrial-superior vena caval junction. Ventricular tachycardia of stable morphology, polymorphic ventricular tachycardia, or ventricular fibrillation was induced four times in each patient and 1, 5, 10, and 25 J truncated exponential shocks with 60% tilt were given in a random sequence. The conversion rate was constant (77%, 86%, 87%, 85%) with increasing energy for ventricular tachycardia but progressively increased for polymorphic ventricular tachycardia and ventricular fibrillation (8%, 33%, 58%, 92%). The ventricular tachycardia acceleration rates for 1, 5, 10, and 25 J were 23%, 14%, 10%, and 15%. Patients not reliably converted with 25 J may require repositioning of leads or two patches. We conclude that for the spring-patch electrodes, increasing energy from 1 to 25 J improves the conversion rate for polymorphic ventricular tachycardia and ventricular fibrillation; the ventricular tachycardia conversion rate is constant. Acceleration of ventricular tachycardia occurs at all energies. Defibrillator implantation requires extensive intraoperative electrophysiologic testing to ensure safe and reliable termination of ventricular tachycardia and fibrillation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Arrhythmia/Electrophysiology Ascending-Ramp Biphasic Waveform Has a Lower Defibrillation Threshold and Releases Less Troponin I Than a Truncated Exponential Biphasic Waveform

Background—We tested the hypothesis that the shape of the shock waveform affects not only the defibrillation threshold but also the amount of cardiac damage. Methods and Results—Defibrillation thresholds were determined for 11 waveforms—3 ascending-ramp waveforms, 3 descending-ramp waveforms, 3 rectilinear first-phase biphasic waveforms, a Gurvich waveform, and a truncated exponential biphasic ...

متن کامل

Ascending-ramp biphasic waveform has a lower defibrillation threshold and releases less troponin I than a truncated exponential biphasic waveform.

BACKGROUND We tested the hypothesis that the shape of the shock waveform affects not only the defibrillation threshold but also the amount of cardiac damage. METHODS AND RESULTS Defibrillation thresholds were determined for 11 waveforms-3 ascending-ramp waveforms, 3 descending-ramp waveforms, 3 rectilinear first-phase biphasic waveforms, a Gurvich waveform, and a truncated exponential biphasi...

متن کامل

Internal cardioversion of atrial fibrillation in sheep.

BACKGROUND The cardioversion efficacy of multiple defibrillation waveforms and electrode systems was compared in a sheep model of atrial fibrillation. METHODS AND RESULTS Sustained atrial fibrillation could be induced with rapid atrial pacing in 23 (55%) of the animals. This study was performed in four parts. Six sheep with sustained atrial fibrillation were used for data analysis for each pa...

متن کامل

Axillary or a Pectoral Thoracic Electrode Location

Background. A significant proportion of patients receiving endocardial defibrillation lead systems must accept either high defibrillation thresholds (DFTs) with lower safety margins or lead implantation by thoracotomy. We examined the feasibility of achieving universal application of endocardial leads and lower defibrillation energy requirements by optimizing the lead system location in conjunc...

متن کامل

Internal Cardioversion of Atrial Fibrillation in Sheep Randolph

Background. The cardioversion efficacy of multiple defibrillation waveforms and electrode systems was compared in a sheep model of atrial fibrillation. Methods and Results. Sustained atrial fibrillation couldbe induced with rapid atrial pacing in 23 (55%) of the animals. This study was performed in four parts. Six sheep with sustained atrial fibrillation were used for data analysis for each par...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 69 4  شماره 

صفحات  -

تاریخ انتشار 1984