Is a quasi-3D dosimeter better than a 2D dosimeter for Tomotherapy delivery quality assurance?
نویسندگان
چکیده
Delivery quality assurance (DQA) has been performed for each Tomotherapy patient either using ArcCHECK or MatriXX Evolution in our clinic since 2012. ArcCHECK is a quasi-3D dosimeter whereas MatriXX is a 2D detector. A review of DQA results was performed for all patients in the last three years, a total of 221 DQA plans. These DQA plans came from 215 patients with a variety of treatment sites including head-neck, pelvis, and chest wall. The acceptable Gamma pass rate in our clinic is over 95% using 3mm and 3% of maximum planned dose with 10% dose threshold. The mean value and standard deviation of Gamma pass rates were 98.2% ± 1.98(1SD) for MatriXX and 98.5%±1.88 (1SD) for ArcCHECK. A paired t-test was also performed for the groups of patients whose DQA was performed with both the ArcCHECK and MatriXX. No statistical dependence was found in terms of the Gamma pass rate for ArcCHECK and MatriXX. The considered 3D and 2D dosimeters have achieved similar results in performing routine patient-specific DQA for patients treated on a TomoTherapy unit.
منابع مشابه
Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems
3D Gamma index is one of the metrics which have been widely used for clinical routine patient specific quality assurance for IMRT, Tomotherapy and VMAT. The algorithms for calculating the 3D Gamma index using global and local methods implemented in two software tools: PTWVeriSoft ® as a part of OCTIVIUS 4D dosimeter systems and 3DVH TM from Sun Nuclear were assessed. The Gamma index calculated ...
متن کاملEvaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA.
The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (<0.1%), dose rate dependence (0.4%), and absolute calibration accuracy (0.4% in the center of the phantom compared with the independently calibrated diode). Relative calibration of ...
متن کاملMeasurement‐guided volumetric dose reconstruction for helical tomotherapy
It was previously demonstrated that dose delivered by a conventional linear accelerator using IMRT or VMAT can be reconstructed - on patient or phantom datasets - using helical diode array measurements and a technique called planned dose perturbation (PDP). This allows meaningful and intuitive analysis of the agreement between the planned and delivered dose, including direct comparison of the d...
متن کاملQuality Assurance / Phantoms
Gamma index pass rates were evaluated for an anthropomorphic phantom using both 2D and 3D calculations. The phantom was irradiated with the traditional dosimetry insert loaded with radiochromic film and TLD, and then with a 3D dosimetry insert. A comparison with the calculated dose distribution showed that both the PRESAGE® dosimeter and the film and TLD system agreed with the plan to within 5%...
متن کاملCharacterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific quality assurance in rapidarc treatment delivery.
A commercial 2D array seven29 detector has been characterized and its performance has been evaluated. 2D array ionization chamber equipped with 729 ionization chambers uniformly arranged in a 27 × 27 matrix with an active area of 27 × 27 cm² was used for the study. An octagon-shaped phantom (Octavius Phantom) with a central cavity is used to insert the 2D ion chamber array. All measurements wer...
متن کامل