Cyclic fatigue resistance of OneShape and WaveOne instruments using different angles of curvature.

نویسندگان

  • Shaymaa E Elsaka
  • Amr M Elnaghy
چکیده

This study compared the cyclic fatigue resistance of OneShape (OS) instrument which used continuous rotation with WaveOne (WO) instrument which used reciprocating motion. OS size 25, 0.06 taper and WO primary size 25, 0.08 taper were used in simulated canals with 45°, 60°, and 90° angles of curvature until fracture. The number of cycles to fracture (NCF) was recorded to evaluate cyclic fatigue resistance. Data were statistically analyzed using two-way analysis of variance and Tukey's HSD test. Weibull analysis was performed on NCF data. WO had a significantly higher NCF than OS for all angles of curvature (p<0.05). According to Weibull distribution, WO instrument was predicted to have a higher number of cycles to survive than OS instrument. WO instrument which used the reciprocating motion had a greater resistance to cyclic fatigue than OS which used continuous rotation movement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of movement kinematics on the cyclic fatigue resistance of nickel-titanium instruments.

AIM To compare the cyclic fatigue resistance of new generation OneShape and WaveOne instruments when used with different kinematics. METHODOLOGY An artificial canal, made of stainless steel with an inner diameter of 1.5 mm, a 60° angle of curvature and a curvature radius of 3 mm, was used for cyclic fatigue test of new generation OneShape and WaveOne instruments. Groups were as follows (n = 2...

متن کامل

Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments

OBJECTIVES To compare the fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments under static model via artificial canals with different angles of curvature. MATERIALS AND METHODS Reciproc R25, WaveOne GOLD Primary, and Tango-Endo instruments were included in this study (n = 20). All the instruments were rotated in artificial canals which were made of stainless steel with an ...

متن کامل

Cyclic fatigue of Reciproc and WaveOne reciprocating instruments.

AIM To evaluate the cyclic fatigue resistance of Reciproc(®) and WaveOne(®) instruments in simulated root canals. METHODOLOGY Two groups of 15 NiTi endodontic instruments of identical tip size of 0.25 mm were tested, group A; Reciproc(®) R25 and group B: WaveOne(®) primary. Cyclic fatigue testing was performed in a stainless steel artificial canal manufactured by reproducing the instrument's ...

متن کامل

Cyclic Fatigue Resistance and Force Generated by OneShape Instruments during Curved Canal Preparation

OBJECTIVES To evaluate the cyclic fatigue resistance and the force generated by OneShape files during preparation of simulated curved canals. METHODS Six OneShape files (the test) and six ProTaper F2 files (the control) were subject to the bending ability test. Another thirty files of each type were used to prepare artificial canals (n = 60), which were divided into 3 groups according to resp...

متن کامل

Differences in cyclic fatigue resistance at apical and coronal levels of Reciproc and WaveOne new files.

INTRODUCTION Recent improvements in alloys, kinematics, and concepts have been combined to increase the cyclic fatigue resistance of nickel-titanium (NiTi) instruments. The aim of this study was to compare the cyclic fatigue resistance of new M-Wire reciprocating WaveOne (Dentsply Maillefer, Ballaigues, Switzerland) and Reciproc (VDW GmbH, Munich, Germany) files at 2 levels. METHODS Sixty Rec...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Dental materials journal

دوره 34 3  شماره 

صفحات  -

تاریخ انتشار 2015