Metal-on-Metal Hip Retrieval Analysis: A Case Report

نویسندگان

  • Thomas B. Pace
  • Kara A. Rusaw
  • Lawrence J. Minette
  • Brayton R. Shirley
  • Rebecca G. Snider
  • John D. DesJardins
چکیده

This is a case report involving a single case with severe bone and soft tissue destruction in a young male patient with a 10-year-metal on-metal total hip arthroplasty. Following complete aseptic erosion of the affected hip greater trochanter and abductor muscles, the hip was revised for recurrent instability. Histological examination of the patient's periprosthetic tissues, serological studies, and review of recent medical reports of similar cases were used to support an explanation of the destructive process and better contribute to our understanding of human reaction to metal debris in some patients following metal-on-metal hip arthroplasty.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Fatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature

Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...

متن کامل

Arthroprosthetic cobaltism associated with cardiomyopathy

Systemic cobaltism related to metal-on-metal total hip arthroplasty has been published in case reports and series with effects on the cardiac, neurologic, endocrine, and immunologic systems. This case report presents a 46-year-old male who underwent bilateral metal-on-metal total hip arthroplasty and subsequently developed cardiomyopathy requiring left ventricular assist device implantation. In...

متن کامل

Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design

OBJECTIVES The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromi...

متن کامل

A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties.

Given the recent reports of high failure rates, an improved understanding of the mechanism of failure of large diameter metal-on-metal hip arthroplasties is essential. We present clinical data and tribological analysis of a consecutive series of 74 failed large diameter metal-on-metal hips, comparing the Durom (Zimmer) with the Birmingham hip resurfacing (BHR) (Smith and Nephew). We retrospecti...

متن کامل

Good function but very high concentrations of cobalt and chromium ions in blood 37 years after metal-on-metal total hip arthroplasy

This case report describes a patient who had a metal-on-metal hip prosthesis implanted 37 years ago. The hip function and X-ray are presented. The levels of cobalt and chromium ions in blood are analyzed and found to be about 40 times higher than normal. Consequences are discussed.

متن کامل

ذخیره در منابع من