Third party brachytherapy seed calibrations and physicist responsibilities.
نویسندگان
چکیده
To the Editor, In the last decade, the number of low-energy photonemitting brachytherapy source manufacturers has dramatically grown, and there are now about 15 manufacturers for I and Pd encapsulated sources for permanent and temporary brachytherapy. More recently, several third party calibration services, some based in commercial radiopharmacies, have begun marketing “independent assays” of brachytherapy source strength. For a fee, these services perform seed strength assay for an order prior to shipping it to the end user. While such services help reduce the physics workload required in source preparation for brachytherapy implants, they also raise certain medical physics, patient safety, and legal issues regarding American Association of Physicists in Medicine AAPM recommendations that have been published in Medical Physics. In 1997, AAPM TG-56 guidance states, “Every institution practicing brachytherapy shall have a system for measuring source strength with secondary traceability for all source types used in its practice.” The report further states, “The institution should compare the manufacturer’s stated value with the institution’s standard.” In 1999, the issue of whether source strength can be verified by entities other than the final user’s institution was addressed again in Medical Physics when AAPM TG-64 restated and condensed TG-56 recommendations as, “In whatever form the seeds are procured, the manufacturer’s assay must be independently confirmed.” In this letter, we intend to update the Medical Physics readership regarding the current deliberations and actions that are underway in addressing these urgent issues raised by the use of third party brachytherapy seed calibrations. This letter has been prepared by members of the AAPM Brachytherapy Subcommittee of the Therapy Physics Committee and was approved for publication by the AAPM Therapy Physics Committee. In our opinion, use of third party calibration services that provide independent source strength verification would appear to provide nominal adherence to TG-64, but not necessarily to TG-56. Additionally, use of such calibrations to replace TG-56 compliant end-user measurements raises questions and concerns. It is the AAPM’s position that a qualified medical physicist is responsible for the dosimetric accuracy of brachytherapy treatment plans, including source strengths. This position is also supported by state regulatory bodies and professional organizations such as the American College of Radiology ACR , American Brachytherapy Society ABS , American College of Medical
منابع مشابه
Third-party brachytherapy source calibrations and physicist responsibilities: report of the AAPM Low Energy Brachytherapy Source Calibration Working Group.
The AAPM Low Energy Brachytherapy Source Calibration Working Group was formed to investigate and recommend quality control and quality assurance procedures for brachytherapy sources prior to clinical use. Compiling and clarifying recommendations established by previous AAPM Task Groups 40, 56, and 64 were among the working group's charges, which also included the role of third-party handlers to...
متن کاملPermanent prostate seed implant brachytherapy: report of the American Association of Physicists in Medicine Task Group No. 64.
There is now considerable evidence to suggest that technical innovations, 3D image-based planning, template guidance, computerized dosimetry analysis and improved quality assurance practice have converged in synergy in modern prostate brachytherapy, which promise to lead to increased tumor control and decreased toxicity. A substantial part of the medical physicist's contribution to this multi-d...
متن کاملAnalysis and practical use: the Abt Study of Medical Physicist Work Values for Radiation Oncology Physics Services--round II.
PURPOSE The initial Abt Study of Medical Physicist Work Values for Radiation Oncology Physics Services was published in October 1995. That study measured qualified medical physicist (QMP) work associated only with routine radiation oncology procedures. In the intervening years, medical physics practice has changed dramatically. Three-dimensional treatment planning, once considered a special pro...
متن کاملTreasure of the Past IX: Exposure Standardization of Iodine-125 Seeds Used for Brachytherapy
A method for calibrating iodine-125 seeds in terms of exposure has been established. The standard free-air ionization chamber, used for measuring soft x rays, was chosen for the measurements. Arrays of four to six seeds were used to enhance the ionization-current-to-background-current ratio. Seeds from an array were measured individually in a re-entrant chamber. The quotient of the exposure rat...
متن کاملA quality management program in intravascular brachytherapy.
While simple, intravascular brachytherapy (IVB) presents a considerable potential for harm to the patient. The medical physicist maintains the responsibility to minimize the likelihood of operational problems or dosimetric errors. The principals for safe operation remain the same as with any radiotherapy treatment: to deliver the correct dose, to the correct location, safety. To develop an effe...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 33 1 شماره
صفحات -
تاریخ انتشار 2006