Accuracy of semen counting chambers as determined by the use of latex beads.

نویسندگان

  • E K Seaman
  • E Goluboff
  • N BarChama
  • H Fisch
چکیده

OBJECTIVE To assess the accuracy of the Hemacytometer (Hausser Scientific, Horsham, PA), Makler (Sefi-Medical Instrument, Haifa, Israel), Cell-VU (Millennium Sciences Inc., New York, NY), and Micro-Cell chambers (Conception Technologies, San Diego, CA) counting chambers. DESIGN A solution containing a known concentration of latex beads was used as the standard to perform counts on the four different counting chambers. MAIN OUTCOME MEASURES Bead counts for the four different chambers were compared with the bead counts of the standard solution. Variability within chambers also was determined. RESULTS Mean bead concentrations for both the Cell-VU and Micro-Cell chambers were consistently similar to the bead concentration of the standard solution. Both the hemacytometer and the Makler chambers overestimated the actual bead concentration of the standard solution by as much as 50% and revealed significant interchamber variability. CONCLUSIONS Our data revealed marked differences in the accuracy and reliability of the different counting chambers tested and emphasized the need for standardization and quality control of laboratory procedures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses

Background Although the depth of the counting chamber is an important factor influencing sperm counting, no research has yet been reported on the measurement and comparison of the depth of the chamber. We measured the exact depths of six kinds of sperm counting chambers and evaluated their accuracy. MaterialsAndMethods In this prospective study, the depths of six kinds of sperm counting chamber...

متن کامل

The use of latex beads in external quality assurance and internal quality control for routine semen analysis.

The usefulness of latex beads of defined concentration was assessed as a substitute for sperm in the performance of External Quality Assurance (EQA) and Internal Quality Control (IQC) of semen analysis. Within the EQA programme, mean±SEM bias (%) was significantly reduced in 2007 compared to 2002 for both specialist (6.0%±5.4% vs. 55.0%±5.9%) and non-specialist (18.4%±5.9% vs. 90.9%±13.4%) labo...

متن کامل

Validation of sperm counting methods using limits of agreement.

A variety of methods exist for counting sperm. Since the introduction of semen analysis, one of these methods, the hemocytometer, has been regarded as the gold standard by andrology laboratories and the World Health Organization (WHO, 1999). The flexible features of this approach, involving fixation and immobilization of sperm, dilution of highly concentrated samples, and the counting of sperm ...

متن کامل

Accuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses.

BACKGROUND Although the depth of the counting chamber is an important factor influencing sperm counting, no research has yet been reported on the measurement and comparison of the depth of the chamber. We measured the exact depths of six kinds of sperm counting chambers and evaluated their accuracy. MATERIALS AND METHODS In this prospective study, the depths of six kinds of sperm counting cha...

متن کامل

Development of a Latex Agglutination Method for Diagnosis of Rotavirus Infections

Background: Rotavirus is a major cause of morbidity and mortality among children with gastroenteritis. Since the discovery of rotaviruses, several techniques have been used for their laboratory diagnosis; those included Electron Microscopy (EM) and enzyme immunoassay. These methods, however, are expensive and not readily available everywhere. We have developed a technique which can be used for ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Fertility and sterility

دوره 66 4  شماره 

صفحات  -

تاریخ انتشار 1996