Clinical comparison of the Spiegelberg parenchymal transducer and ventricular fluid pressure.
نویسندگان
چکیده
The Spiegelberg brain pressure catheter is a low cost implantable intracranial pressure measuring system which has the unique ability to perform regular automatic zeroing. A new version of the catheter has become available with a subdural bolt fixation to allow insertion of the device into the brain parenchyma. The accuracy of this system has been evaluated in comparison with a ventricular fluid pressure method in a series of patients to determine its accuracy and utility in the clinical environment. Hourly readings from the Spiegelberg system have been compared with those obtained using a standard pressure transducer connected to an external ventricular drain. Measurements continued while there was a clinical need for CSF drainage. Eleven patients were recruited to the study and data were recorded for periods ranging from 40 to 111 hours. A good agreement between the two systems was obtained. In 10 cases the mean difference was less than +/-1.5 mm Hg and the dynamic changes in value were contemporaneous. In one case an intracerebral haemorrhage developed around the tips of the Spiegelberg catheter and significant differences occurred between the two methods of measurement. In conclusion, the Spiegelberg parenchymal transducer provides an accurate measurement of intracranial pressure when compared with ventricular pressure. The transducer was found to be robust in the clinical environment and very popular with the nursing staff. Further studies may determine whether the complication rate of this system is comparable with other available devices.
منابع مشابه
Laboratory testing of the Spiegelberg brain pressure monitor: a technical report.
OBJECTIVES The Spiegelberg brain pressure monitor is a low cost intracranial pressure monitoring system that has been used clinically for some time, mainly in Germany. To provide a rigorous bench comparison of the Spiegelberg monitor with the Camino pressure monitor an evaluation programme has been carried out in the UK Shunt Evaluation Laboratory. DESIGN Drift over 72 hours and with temperat...
متن کاملExperimental evaluation of the Spiegelberg intracranial pressure and intracranial compliance monitor. Technical note.
The goal of this study was to compare the Spiegelberg intraventricular intracranial pressure (ICP)/intracranial compliance monitoring device, which features an air-pouch balloon catheter, with existing gold-standard methods of measuring ICP and intracranial compliance. A Spiegelberg intraventricular catheter, a standard intraventricular catheter, and a Codman intraparenchymal ICP microsensor we...
متن کاملComparison of parenchymal and ventricular intracranial pressure readings utilizing a novel multi-parameter intracranial access system
INTRODUCTION Both ventricular and parenchymal devices are available for measurement of intracranial pressure (ICP). The Hummingbird(®) Synergy Ventricular System is a novel device allowing multi-parametric neurological monitoring, including both ventricular and parenchymal ICP. The purpose of this study is to compare the congruence of the device's ventricular and parenchymal ICP readings. MET...
متن کاملCerebrospinal Pulsation Hydrodynamics in a 2D Simulation of Brain Ventricles
In this article, dynamics of the cerebrospinal fluid (CSF) was studied, using computational fluid dynamics. Using MRI images of two special cases, a 2-dimensional model of the ventricular system was made. CSF velocity and pressure distribution in ventricular system have high importance since the flow pattern of this liquid has an important effect on intracranial pressure, i.e., ICP, which has a...
متن کامل2D Computational Fluid Dynamic Modeling of Human Ventricle System Based on Fluid-Solid Interaction and Pulsatile Flow
Many diseases are related to cerebrospinal .uid (CSF) hydrodynamics. Therefore, understanding the hydrodynamics of CSF .ow and intracranial pressure is helpful for obtaining deeper knowledge of pathological processes and providing better treatments. Furthermore, engineering a reliable computational method is promising approach for fabricating in vitro models which is essential for inventing gen...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of neurology, neurosurgery, and psychiatry
دوره 71 3 شماره
صفحات -
تاریخ انتشار 2001