Comparative study of six rotary nickel-titanium systems and hand instrumentation for root canal preparation.

نویسندگان

  • A Guelzow
  • O Stamm
  • P Martus
  • A M Kielbassa
چکیده

AIM To compare ex vivo various parameters of root canal preparation using a manual technique and six different rotary nickel-titanium (Ni-Ti) instruments (FlexMaster, System GT, HERO 642, K3, ProTaper, and RaCe). METHODOLOGY A total of 147 extracted mandibular molars were divided into seven groups (n = 21) with equal mean mesio-buccal root canal curvatures (up to 70 degrees), and embedded in a muffle system. All root canals were prepared to size 30 using a crown-down preparation technique for the rotary nickel-titanium instruments and a standardized preparation (using reamers and Hedströem files) for the manual technique. Length modifications and straightening were determined by standardized radiography and a computer-aided difference measurement for every instrument system. Post-operative cross-sections were evaluated by light-microscopic investigation and photographic documentation. Procedural errors, working time and time for instrumentation were recorded. The data were analysed statistically using the Kruskal-Wallis test and the Mann-Whitney U-test. RESULTS No significant differences were detected between the rotary Ni-Ti instruments for alteration of working length. All Ni-Ti systems maintained the original curvature well, with minor mean degrees of straightening ranging from 0.45 degrees (System GT) to 1.17 degrees (ProTaper). ProTaper had the lowest numbers of irregular post-operative root canal diameters; the results were comparable between the other systems. Instrument fractures occurred with ProTaper in three root canals, whilst preparation with System GT, HERO 642, K3 and the manual technique resulted in one fracture each. Ni-Ti instruments prepared canals more rapidly than the manual technique. The shortest time for instrumentation was achieved with System GT (11.7 s). CONCLUSIONS Under the conditions of this ex vivo study all Ni-Ti systems maintained the canal curvature, were associated with few instrument fractures and were more rapid than a standardized manual technique. ProTaper instruments created more regular canal diameters.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بررسی مشکلات آماده سازی کانال با وسائل دستی و چرخشی نیکل تیتانیوم

Background and Aim: Maintaining the original canal path and safety of canal preparation during instrumentation in curved canals have been subjected to many challenges. The aim of this study was to compare several parameters of root canal preparation using flexmaster rotary and hand used Nickel-Titanium instruments. Materials and Methods: In this experimental study, sixty extracted human teeth w...

متن کامل

A comparison of stainless steel hand and rotary nickel-titanium instrumentation using a silicone impression technique.

BACKGROUND Root canal preparation using rotary nickel-titanium instruments has been reported to be superior to hand instrumentation in terms of root canal shape, centring and frequency of procedural errors. The purpose of this study was to evaluate canal preparations using a sequential silicone impression technique to assess root canal morphology. METHODS Pre-operative canal impressions were ...

متن کامل

Chemomechanical preparation by hand instrumentation and by Mtwo engine-driven rotary files, an ex vivo study

OBJECTIVE To compare the disinfecting efficacy of the sodium hypochlorite irrigation by root canal preparation with stainless steel hand files, taper 0.02 and nickel-titanium Mtwo files with taper 0.04-0.06. STUDY DESIGN 40 extracted human teeth were sterilized, and then inoculated with Enterococcus faecalis (ATCC 29212). After 6 day incubation time the root canals were prepared by hand with ...

متن کامل

Comparison of working length control consistency between hand K-files and Mtwo NiTi rotary system

BACKGROUND The purpose of this study was to investigate the consistency of working length control between hand instrumentation in comparison to engine driven Mtwo nickel-titanium rotary files. MATERIAL AND METHODS Forty extracted maxillary molars were selected and divided onto two parallel groups. The working lengths of the mesiobuccal root canals were estimated. The teeth were fixed in a pha...

متن کامل

Comparative study on canal preparation by three rotary sys-tems using CT scan (An in vitro study)

Comparative study on canal preparation by three rotary sys-tems using CT scan (An in vitro study) Dr. M.H. Zarrabi* - Dr. A. Talati** - Dr. S. Moradi** - Dr. L. Shakeri*** * - Associate Professor and Chief of Endodontics Dept. - Faculty of Dentistry – Mashhad University of Medical Sciences. **- Assistant Professor of Endodontics Dept. - Faculty of Dentistry – Mashhad University of Medical Scien...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International endodontic journal

دوره 38 10  شماره 

صفحات  -

تاریخ انتشار 2005