Transthoracic defibrillation: effect of sternotomy on chest impedance.

نویسندگان

  • R E Kerber
  • S Vance
  • S J Schomer
  • D J Mariano
  • F Charbonnier
چکیده

The purpose of this study was to determine the effect of sternotomy on transthoracic impedance, a major determinant of current flow and defibrillation success. Transthoracic impedance was determined by using a validated test-pulse technique that does not require actual shocks. Seventeen patients undergoing median sternotomy were studied prospectively. Transthoracic impedance was determined before operation, 3 to 5 days after operation and (in eight patients) greater than or equal to 1 month after operation. When measured using paddle electrodes placed in the standard apex-right parasternal defibrillating position, transthoracic impedance declined after sternotomy in all patients, from 77 +/- 18 to 59 +/- 17 omega (p less than 0.01); smaller declines were demonstrated by using other electrode positions. Transthoracic impedance remained below the preoperative level in the eight patients who underwent a second set of measurements at least 1 month after operation. Six normal subjects not undergoing sternotomy underwent serial transthoracic impedance measurements at least 5 days apart; mean transthoracic impedance did not change. It is concluded that transthoracic impedance declines after sternotomy. At any operator-selected energy level a higher current flow will result after sternotomy; this may facilitate postoperative defibrillation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

defibrillation: experimental studies

In defibrillation, current flow depends on the energy selected and the transthoracic impedance. If transthoracic impedance is high, current flow may be inadequate to defibrillate. We developed a method by which high transthoracic impedance is automatically compensated for by an increase in operator-selected energy when impedance is high. Transthoracic impedance was predicted in advance of the f...

متن کامل

Self-adhesive preapplied electrode pads for defibrillation and cardioversion.

The efficacy of self-adhesive electrode pads for defibrillation and cardioversion was assessed in 80 patients who received 267 shocks from self-adhesive pads. In all but two patients, defibrillation or cardioversion was achieved at least once. The pads were equally effective when used in the apex-anterior or apex-posterior position. The transthoracic impedance using self-adhesive pads was 75 +/...

متن کامل

BRIEF REVIEW Determinants of successful defibrillation

Ventricular defibrillation by electric counter shock is the definitive treatment for cardiac arrest due to ventricular fibrillation (VF). Ventricular defibrillation is the fundamental cornerstone to a successful outcome. As most deaths from VF occur out of hospital, often in patients without pre-existent recognised cardiac disease, the major public health issue remains the widespread applicatio...

متن کامل

WORKSHEET for Evidence-Based Review of Science for Emergency Cardiac Care Worksheet author(s)

Search strategy (including electronic databases searched). DRAJER: Several search strategies were used for each data base. Only those with maximal hits were considered Embase “defibrillation” AND “pads” AND “impedance” as text words (all fields): 31 hits Embase “defibrillation” AND “pads” AND [<1966-2010]/py: 85 hits PubMed “defibrillation” AND “electrode pads” as text words (all fields): 52 hi...

متن کامل

Transthoracic impedance does not affect defibrillation, resuscitation or survival in patients with out-of-hospital cardiac arrest treated with a non-escalating biphasic waveform defibrillator.

OBJECTIVE This is a study of the influence of transthoracic impedance (TTI) on defibrillation, resuscitation and survival in patients with out-of-hospital cardiac arrest (OHCA), treated with a non-escalating impedance-compensating 150 J biphasic waveform defibrillator. METHODS Cardiac arrest data from two EMS systems were analyzed retrospectively. All witnessed arrests from patients who prese...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 20 1  شماره 

صفحات  -

تاریخ انتشار 1992