Verification of in vitro medical diagnostics (IVD) metrological traceability: responsibilities and strategies.
نویسندگان
چکیده
To be accurate and equivalent, laboratory results should be traceable to higher-order references. Furthermore, their analytical performance should fulfill acceptable measurement uncertainty criteria defined to fit the intended clinical use. With this aim, In Vitro Diagnostics (IVD) manufacturers should define a calibration hierarchy to assign traceable values to their system calibrators and to fulfill during this process uncertainty limits for calibrators, which should represent a proportion of the uncertainty budget allowed for laboratory results. It is important that end-users may know and verify how manufacturers have implemented the traceability of their calibrators and estimated the corresponding uncertainty. However, full information about traceability and combined uncertainty of calibrators is currently not available. Important tools for IVD traceability surveillance are the verification by laboratories of the consistency of declared performance during daily operations performed in accordance with the manufacturer's instructions and the organization of appropriately structured External Quality Assessment (EQA) programs. The former activity should be accomplished by analyzing system control materials and confirming that current measurements are in the manufacturer's established control range. With regard to EQA, it is mandatory that target values for materials are assigned with reference procedures by accredited laboratories, that materials are commutable and that a clinically allowable inaccuracy for participant's results is defined.
منابع مشابه
Role and Responsibilities of Laboratory Medicine Specialists in the Verification OF Metrological Traceability of in vitro Medical Diagnostics
To be accurate and equivalent, laboratory results should be traceable to higher-order references. Furthermore, their quality should fulfill acceptable measurement uncertainty as defined to fit the intended clinical use. With this aim, in vitro diagnostics (IVD) manufacturers should define a calibration hierarchy to assign traceable values to their system calibrators and to fulfill during this p...
متن کاملPanteghini Performance criteria for combined uncertainty budget in the implementation of metrological traceability
The measurement uncertainty budget should combine the uncertainty of higher order references, the uncertainty of commercial system calibration, the system imprecision and individual laboratory performance in terms of variability. Here we recommend that no more than one third of the total uncertainty budget, established by appropriate analytical performance specifications, is consumed by the unc...
متن کاملPanteghini Performance criteria for combined uncertainty budget in the implementation of metrological
The measurement uncertainty budget should combine the uncertainty of higher order references, the uncertainty of commercial system calibration, the system imprecision and individual laboratory performance in terms of variability. Here we recommend that no more than one third of the total uncertainty budget, established by appropriate analytical performance specifications, is consumed by the unc...
متن کاملPerformance criteria for combined uncertainty budget in the implementation of metrological traceability.
The measurement uncertainty budget should combine the uncertainty of higher order references, the uncertainty of commercial system calibration, the system imprecision and individual laboratory performance in terms of variability. Here we recommend that no more than one third of the total uncertainty budget, established by appropriate analytical performance specifications, is consumed by the unc...
متن کاملHarmonization of Clinical Laboratory Test Results: The Role of the IVD Industry
At the start of the 21st century, a dramatic change occurred in the clinical laboratory community. Concepts from Metrology, the science of measurement, began to be more carefully applied to the in vitro diagnostic (IVD) community, that is, manufacturers. A new appreciation of calibrator traceability evolved. Although metrological traceability always existed, it was less detailed and formal. The...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinica chimica acta; international journal of clinical chemistry
دوره 432 شماره
صفحات -
تاریخ انتشار 2014