Evaluation of an X-Ray Dose Profile Derived from an Optically Stimulated Luminescent Dosimeter during Computed Tomographic Fluoroscopy
نویسندگان
چکیده
The purpose of this study was to evaluate scatter radiation dose to the subject surface during X-ray computed tomography (CT) fluoroscopy using the integrated dose ratio (IDR) of an X-ray dose profile derived from an optically stimulated luminescent (OSL) dosimeter. We aimed to obtain quantitative evidence supporting the radiation protection methods used during previous CT fluoroscopy. A multislice CT scanner was used to perform this study. OSL dosimeters were placed on the top and the lateral side of the chest phantom so that the longitudinal direction of dosimeters was parallel to the orthogonal axis-to-slice plane for measurement of dose profiles in CT fluoroscopy. Measurement of fluoroscopic conditions was performed at 120 kVp and 80 kVp. Scatter radiation dose was evaluated by calculating the integrated dose determined by OSL dosimetry. The overall percent difference of the integrated doses between OSL dosimeters and ionization chamber was 5.92%. The ratio of the integrated dose of a 100-mm length area to its tails (-50 to -6 mm, 50 to 6 mm) was the lowest on the lateral side at 80 kVp and the highest on the top at 120 kVp. The IDRs for different measurement positions were larger at 120 kVp than at 80 kVp. Similarly, the IDRs for the tube voltage between the primary X-ray beam and scatter radiation was larger on the lateral side than on the top of the phantom. IDR evaluation suggested that the scatter radiation dose has a high dependence on the position and a low dependence on tube voltage relative to the primary X-ray beam for constant dose rate fluoroscopic conditions. These results provided quantitative evidence supporting the radiation protection methods used during CT fluoroscopy in previous studies.
منابع مشابه
Evaluation of the Quality Control Program for Diagnostic Radiography and Fluoroscopy Devices in Syria during 2005-2013
Introduction: Extensive use of diagnostic radiology is the largest contributor to total population radiation doses. Thus, appropriate equipment and safe practice are necessary for good-quality images with optimal doses. This study aimed to perform quality control (QC) audit for radiography and fluoroscopy devices owned by private sector in Syria (2005-2013) to verify compliance of performance o...
متن کاملOrgan Dose Measurement in Computed Tomography Using Thermoluminescence Dosimeter in Locally Developed Phantoms
Introduction: Organ dose estimation using thermoluminescence dosimeter (TLD) is known to be a standard, although many other methods, such as simulation software, optically stimulated luminescent dosimeters, and photodiodes are still in use. This study aimed at directly measuring mean organ doses to the selected organs in the head/neck, chest, and abdominal regions from four computed tomography ...
متن کاملEstimation of the effective dose to the radiologists during fluoroscopy or angiography of abdominal viscera
ABSTRACTBackground: as fluoroscopy or different angiographies are usually evaluated using TL dosimeter. Thismethod is a time-consuming procedure therefore, radiologists are usually worried and unawareof their exact radiation dose received during each fluoroscopy or angiography. In this study, anew trend for quick estimation of effective dose based on measuring air-absorbed dose of thescattered ...
متن کاملEvaluation of Maximum Patient Skin Dose Arising from Interventional Cardiology Using Thermoluminescence Dosimeter in Mashhad, Iran
Introduction The increasing practice of interventional fluoroscopy in diagnosis and treatment of cardiovascular disease has risen attention to improve radiation protection of patients and cardiologists in these relatively high dose techniques. Therefore, nowadays there is an emphasis on the measurement of radiation dose received by patients and cardiologists arising from the relevant procedures...
متن کاملMeasurement of skin dose from cone-beam computed tomography imaging
OBJECTIVE To measure surface skin dose from various cone-beam computed tomography (CBCT) scanners using point-dosimeters. MATERIALS & METHODS A head anthropomorphic phantom was used with nanoDOT optically stimulated luminescence (OSL) dosimeters (Landauer Corp., Glenwood, IL) attached to various anatomic landmarks. The phantom was scanned using multiple exposure protocols for craniofacial eva...
متن کامل