The Shear Strength of the VLPTM Locking Osteopenia Screw versus the VLP Locking Cortical Screw in a PERI-LOCTM VLP One-Third Locking Tubular Plate Construct
نویسنده
چکیده
Deformity correction and internal fixation during surgical treatment of Charcot Arthropathy (CA) requires strong implants capable of ensuring long-term stability and function of the foot. The purpose of this study was to determine the construct shear strength of the newly developed VLP Locking Osteopenia Screw (Smith & Nephew, Inc., Memphis, TN, USA), as compared to the VLP Locking Cortical Screw (Smith & Nephew, Inc.) in a PERI-LOC VLP One-Third Locking Tubular Plate (Smith & Nephew, Inc.). A foam bone analog was cut and placed into a testing device to simulate supported and unsupported fracture conditions. The construct was loaded perpendicular to the plate to reproduce the plantar loading patterns of a typical neuropathic CA patient. The simulated fracture was reduced with a PERI-LOC VLP plate and either the osteopenia or cortical locking screws and then loaded until failure. Testing results demonstrated that construct shear strength was approximately 1.7 times higher for the osteopenia screw construct (p<0.05). This evidence suggests that performance of the VLP Locking Osteopenia Screw can be expected to exceed that of the VLP Locking Cortical Screw under similar loading conditions.
منابع مشابه
First Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw
Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...
متن کاملA Pitfall in Fixation of Distal Humeral Fractures with Pre-Contoured Locking Compression Plate
Anatomically precontoured locking plates are intended to facilitate the fixation of articular fractures and particularly those associated with osteoporosis. Fractures of the distal humerus are relatively uncommon injuries where operative intervention can be exceptionally challenging. The distal humeral trochlea provides a very narrow anatomical window through which to pass a fixed-angle locking...
متن کاملFirst Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw
Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...
متن کاملBiomechanical comparison of axial load between cannulated locking screws and noncannulated cortical locking screws.
The goal of this study was to compare the biomechanical stability of cannulated locking screws and noncannulated cortical locking screws in a periarticular locking plate. Twelve fresh-frozen porcine tibias with a 1-cm gap created distal to the tibial plateau were used to simulate an unstable proximal tibial fracture. All specimens were fixed with a periarticular proximal lateral tibial locking ...
متن کاملA biomechanical study of conventional acetabular internal fracture fixation versus locking plate fixation.
BACKGROUND Conventional internal fixation entails the use of an interfragmentary lag screw along with a plate. Not all acetabular fractures are amenable to the placement of an interfragmentary lag screw, and the fracture may be displaced during tightening of the interfragmentary lag screw. Locking plates are a possible solution. We sought to determine whether a locking plate construct can provi...
متن کامل