A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT.
نویسندگان
چکیده
X-ray scatter poses a significant limitation to image quality in cone-beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy. We report the performance of a simple scatter correction method in which scatter fluence is estimated directly in each projection from pixel values near the edge of the detector behind the collimator leaves. The algorithm operates on the simple assumption that signal in the collimator shadow is attributable to x-ray scatter, and the 2D scatter fluence is estimated by interpolating between pixel values measured along the top and bottom edges of the detector behind the collimator leaves. The resulting scatter fluence estimate is subtracted from each projection to yield an estimate of the primary-only images for CBCT reconstruction. Performance was investigated in phantom experiments on an experimental CBCT bench-top, and the effect on image quality was demonstrated in patient images (head, abdomen, and pelvis sites) obtained on a preclinical system for CBCT-guided radiation therapy. The algorithm provides significant reduction in scatter artifacts without compromise in contrast-to-noise ratio (CNR). For example, in a head phantom, cupping artifact was essentially eliminated, CT number accuracy was restored to within 3%, and CNR (breast-to-water) was improved by up to 50%. Similarly in a body phantom, cupping artifact was reduced by at least a factor of 2 without loss in CNR. Patient images demonstrate significantly increased uniformity, accuracy, and contrast, with an overall improvement in image quality in all sites investigated. Qualitative evaluation illustrates that soft-tissue structures that are otherwise undetectable are clearly delineated in scatter-corrected reconstructions. Since scatter is estimated directly in each projection, the algorithm is robust with respect to system geometry, patient size and heterogeneity, patient motion, etc. Operating without prior information, analytical modeling, or Monte Carlo, the technique is easily incorporated as a preprocessing step in CBCT reconstruction to provide significant scatter reduction.
منابع مشابه
Refinement of Imaging Processing of Scatter Correction and Beam Hardening Tools for Industrial Radiography and Cone Beam CT
Digital detector arrays are used for industrial radiography and cone beam CT. Scattering and beam hardening artifacts reduce the achievable contrast sensitivity and can severely limit the size of the part to be scanned or require large amounts of shielding to block scatter outside the inspection area of interest. Varian’s image processing toolkit contains a set of dll’s that can easily be deplo...
متن کاملX-Ray Scatter Correction on Soft Tissue Images for Portable Cone Beam CT.
Soft tissue images from portable cone beam computed tomography (CBCT) scanners can be used for diagnosis and detection of tumor, cancer, intracerebral hemorrhage, and so forth. Due to large field of view, X-ray scattering which is the main cause of artifacts degrades image quality, such as cupping artifacts, CT number inaccuracy, and low contrast, especially on soft tissue images. In this work,...
متن کاملEvaluation of a Physical Based Approach of Scattered Radiation Correction in Cone Beam CT for Non-Destructive Testing Applications
Cone Beam Computerized Tomography (CBCT) enables threedimensional imaging with isotropic resolution. X-ray scatter management is a challenging task for quantitative CBCT imaging : scattered radiation level is significantly high on cone beam systems compared to collimated fan beam systems. The effects of this scattered radiation are cupping artifacts, streaks, and quantification inaccuracies. At...
متن کاملBeam Attenuation Grid–Based Scatter Correction Algorithm for Cone Beam Volume CT
In the flat panel detector-based cone beam CT system, the scatter photons which reach the flat panel detector increase the detected signal, produce the image cupping artifact and reduce the image contrast. To conquer the X-Ray scatter problem, this study proposes the beam attenuation grid -based scatter correction algorithm that needs to add a beam attenuation grid (BAG) between the X-Ray sourc...
متن کاملبررسی مقایسهای دقت Cone beam CT، رادیوگرافی داخل دهانی و پروب پریودنتال در اندازهگیری ضایعات استخوانی پریودنتال
Background and Aims: Cone beam computed tomography (CBCT) produces high-quality data about diagnosis and periodontal treatment. To date, there is not enough research regarding periodontal bone measurement using CBCT. The aim of this study was to compare the accuracy of CBCT in measuring periodontal defects to that of intraoral radiography and probing methods.Materials and Methods: Two-hundred a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 33 1 شماره
صفحات -
تاریخ انتشار 2006