Braided Polyester Suture Fatigue Failure When Treating Transverse Patellar Fracture: A Case Report
نویسندگان
چکیده
The incidence of patellar fractures accounts for almost 1% of all skeletal fractures. In case of a displaced patellar fracture, surgical intervention is often advocated for earlier stability, restoration of knee motion, and shortened morbidity. Among various fixations, the AO modified tension band technique, which incorporates two parallel Kirschner wires with an anterior tension band, is the most widely applied method. In current practice, the stainless steel wire is the material of choice for treatment of these fractures because of its high tensile strength. However, certain concerns about the difficult manipulation of the wire, frequent irritation after implantation, and even later relieving removal prompted investigators to seek alternatives. Some surgeons began using a braided polyester suture as substitute for stainless steel because it evokes minimal tissue reactivity [4] and works as well mechanically as the stainless steel wire. In the literature, failure of non-absorbable suture fixation is seldom reported. In 2001, Gosal et al, reporting their experience using nonmetallic Braided Polyester Suture Fatigue Failure When Treating Transverse Patellar Fracture: A Case Report Cheng-Li Lin 2, , Chih-Wei Chang 2, , Chih-Hang Chang, Chyun-Yu Yang 1 Department of Orthopedics, National Cheng Kung University Hospital, Tainan, Taiwan 2 Department of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan 3 Department of Orthopedics, Show-Chwan Memorial Hospital, Changhua, Taiwan 4 Division of Traumatology, National Cheng Kung University Hospital, Tainan, Taiwan
منابع مشابه
Patellar stress fracture after transosseous extensor mechanism repair: report of 3 cases.
Extensor mechanism injuries are unusual and require early operative intervention to optimally restore extensor mechanism continuity, strength, and function. Quadriceps tendon ruptures typically occur in patients older than 40 years and can be associated with degenerative tendon changes, obesity, systemic illnesses, and trauma. Patellar tendon ruptures are usually seen in younger patients as a r...
متن کاملCyclic testing of pullout sutures and micro-mitek suture anchors in flexor digitorum profundus tendon distal fixation.
PURPOSE Little data exist comparing the strength of traditional methods of fixation in a flexor digitorum profundus tendon with the use of a suture anchor. In vitro cyclic testing simulating a passive mobilization protocol was used to compare the repair of a flexor digitorum profundus tendon using a single micro-Mitek anchor (Mitek, Westwood, MA) or a modified Bunnell 2-strand pullout technique...
متن کاملThe influence of suture material on the strength of horizontal mattress suture configuration for meniscus repair.
PURPOSE Comparison of the mechanical characteristics of meniscal repair fixation using horizontal sutures and six different sutures under submaximal cyclic and load to failure test conditions may aid physicians in selecting a suture type. METHODS A 2-cm long anteroposterior vertical longitudinal incision was created in six groups of bovine medial menisci. Lesions were repaired using a No. 2 s...
متن کاملSerial Radiographs Showing Progression of a Patellar Stress Fracture and Beneficial Surgical Technique for a Displaced Patellar Stress Fracture
Stress fractures of the patella occurring in athletes are quite rare and only 23 cases have been introduced in English literatures. Stress fractures of the patella are classified into two types: transverse and longitudinal. They are mostly seen in the distal one-third of the patella on a plain lateral radiograph. We present here a 16-year-old female basketball player with a stress fracture of t...
متن کاملSuture biomechanics and static facial suspension.
BACKGROUND Static facial suspension (SFS) continues to play a role for rehabilitation in patients with facial paralysis. We perform SFS almost exclusively with a suture technique in our practice. Monofilament polypropylene suture (Prolene) is commonly used for SFS, but we have witnessed occasional failure and some stretching with this material. The purpose of this study was to establish and com...
متن کامل