Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: safety, tolerability, and reproducibility in patients with glaucoma.
نویسندگان
چکیده
OBJECTIVE To examine the safety, tolerability, and reproducibility of intraocular pressure (IOP) patterns during repeated continuous 24-hour IOP monitoring with a contact lens sensor. METHODS Forty patients suspected of having glaucoma(n=21) or with established glaucoma (n=19) were studied.Patients participated in two 24-hour IOP monitoring sessions (S1 and S2) at a 1-week interval (SENSIMED Triggerfish CLS; Sensimed AG). Patients pursued daily activities,and sleep behavior was not controlled. Incidence of adverse events and tolerability (visual analog scale score)were assessed. Reproducibility of signal patterns was assessed using Pearson correlations. RESULTS The mean (SD) age of the patients was 55.5(15.7) years, and 60% were male. Main adverse events were blurred vision (82%), conjunctival hyperemia (80%),and superficial punctate keratitis (15%). The mean (SD)visual analog scale score was 27.2(18.5) mm in S1 and 23.8(18.7) mm in S2 (P=.22). Overall correlation between the 2 sessions was 0.59 (0.51 for no glaucoma medication and 0.63 for glaucoma medication) (P=.12). Mean(SD) positive linear slopes of the sensor signal from wake to 2 hours into sleep were detected in both sessions for the no glaucoma medication group (S1: 0.40 [0.34],P.001; S2: 0.33[0.30], P.01) but not for the glaucoma medication group (S1: 0.24 [0.60], P=.06; S2:0.40[0.40], P.001). CONCLUSIONS Repeated use of the contact lens sensor demonstrated good safety and tolerability. The recorded IOP patterns showed fair to good reproducibility,suggesting that data from continuous 24-hour IOP monitoring may be useful in the management of patients with glaucoma. TRIAL REGISTRATION clinicaltrials.gov Identifier: NCT01319617
منابع مشابه
Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens.
PURPOSE To investigate tolerability and safety of a new diagnostic device for 24-hour intraocular pressure monitoring in healthy subjects and age-matched glaucoma patients. PATIENTS AND METHODS Twenty healthy subjects (group 1) and 20 age-matched glaucoma patients (group 2) were included in this prospective, single-center, open, observational parallel group study. The SENSIMED Triggerfish Sen...
متن کاملTwenty-four-hour intraocular pressure patterns in a symptomatic patient after ab interno trabeculotomy surgery
We report the results of repeated ambulatory continuous 24-hour intraocular pressure (IOP) monitoring with a contact lens sensor (CLS) in a glaucoma patient with ocular pain after ab interno trabeculotomy (Trabectome™) surgery. Our findings show that a combined prostaglandin-pilocarpine treatment reduced nighttime IOP peaks and relieved the patient's symptoms.
متن کاملThe Application of a Contact Lens Sensor in Detecting 24-Hour Intraocular Pressure-Related Patterns
Glaucoma is one of the leading causes of blindness worldwide. Recent studies suggest that intraocular pressure (IOP) fluctuations, peaks, and rhythm are important factors in disease advancement. Yet, current glaucoma management remains hinged on single IOP measurements during clinic hours. To overcome this limitation, 24-hour IOP monitoring devices have been employed and include self-tonometry,...
متن کاملThe Fluctuation of Intraocular Pressure Measured by a Contact Lens Sensor in Normal-Tension Glaucoma Patients and Nonglaucoma Subjects.
PURPOSE We compared the fluctuation of intraocular pressure (IOP) in normal-tension glaucoma (NTG) patients and individuals with nonglaucoma eyes. We obtained continuous IOP values using a SENSIMED Triggerfish contact lens sensor. MATERIALS AND METHODS The eyes of 12 nonglaucoma subjects and 14 NTG patients were examined. In all 26 subjects, the IOP fluctuation was measured continuously for 2...
متن کاملThe Road Ahead to Continuous 24-Hour Intraocular Pressure Monitoring in Glaucoma
Intraocular pressure (IOP) is the only treatable risk factor for glaucoma. Yet, current glaucoma management usually relies on single IOP measurements during clinic hours despite the fact that IOP is a dynamic parameter with individual rhythms. Single IOP measurements underpin all major clinical guidelines on treatment of glaucoma. Other potentially informative parameters, such as IOP fluctuatio...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Archives of ophthalmology
دوره 130 12 شماره
صفحات -
تاریخ انتشار 2012