An Absorbed Dose Water Calorimeter: Theory, Design, and Performance
نویسنده
چکیده
Two calibrated thermistors sandwiched between two insulative polyethylene films were immersed in a 30 cm cube of water having a low thermal diffusivity. The product of the specific heat of water and temperature rise gave the combined effect of the local absorbed dose and any heat defect at a position along a temperature profile produced by cobalt-60 irradiation. The dose rate was near 18 mGy/s and exposure times were 3 min. The standard deviation for a daily set of measurements was about 0.6 percent. Calculations showed that conductive heat transfer produced a negligible effect at the position of measurement along the beam axis. Tests showed the absence of convection. Temperature drifts before irradiation were quickly controlled by changing the power dissipated in the water between two immersed electrodes. Reproducible measurements were obtained in distilled water supplies that had a wide range of impurities. Measurements, after saturating the water with nitrogen or oxygen, showed no difference. A difference of 0.6 percent would have been easily detectable. Tests with several chemicals added to water showed some unexpected results and changes in the measured absorbed dose rate versus accumulated dose. The measured absorbed dose rate in distilled water under the conditions described was 3.5 percent higher than that determined from measurements with a graphite calorimeter.
منابع مشابه
The UK primary standard calorimeter for photon-beam absorbed dose measurement.
A description is given of the UK primary standard graphite calorimeter system. The calorimeter measures absorbed dose to graphite for photon radiations from 60Co to 19 MV x-rays, and is the basis of the NPL therapy-level absorbed dose to water calibration service. Absorbed dose to graphite from the photon calorimeter has been compared with three other standards: an ionization chamber and cavity...
متن کاملDevelopment of a new in-water-phantom graphite calorimeter for the measurement of absorbed dose to water in medium energy x-ray beams
This paper summarizes the progress on the construction of a graphite calorimeter for the measurement of absorbed dose to water Dw in medium-energy filtered x-ray qualities, as part of the EMRPfunded project “Metrology for radiotherapy using complex radiation fields” (MetrExtRT, http://radiotherapyemrp.eu/). The design and the construction of a new primary standard for absorbed dose to water (Dw...
متن کاملA Sealed Water Calorimeter for Measuring Absorbed Dose
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001 The NIST sealed water calorimeter is intended for direct measurement of absorbed dose to water. This calorimeter was used for a series of approximately 3700 measurements to investigate the so-called heat defect, that is, anomalous endothermic or exothermic effects caused by dissolved gases. The three systems investigate...
متن کاملA Polystyrene-Water Calorimeter
The paper describes a new type of calorimeter that can be quickly put into operation for determining absorbed dose at a point in polystyrene. It also describes a unique method of decreasing drifts in electrical signals caused by temperature gradients. Two calibrated thermistors were placed close together between sandwiched polystyrene discs that were immersed in water. The assembly was irradiat...
متن کاملCalorimetry for Absorbed-dose Measurements at BNM-LNHB
At BNM-LNHB, the references in terms of absorbed dose to water are based on graphite calorimeter measurements. In the cobalt-60 beam, measurements were performed directly with the graphite calorimeter in a graphite phantom. Absorbed dose to water in a water phantom was then determined using a transfer method by means of relative measurements made with ionisation chambers and Fricke dosimeters. ...