Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor
نویسندگان
چکیده
BACKGROUND This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (Guardian™ Sensor 3) in the abdomen and arm. METHODS Eighty-eight subjects (14-75 years of age, mean ± standard deviation [SD] of 42.0 ± 19.1 years) with type 1 or type 2 diabetes participated in the study. Subjects wore two sensors in the abdomen that were paired with either a MiniMed™ 640G insulin pump, or an iPhone® or iPod® touch® running a glucose monitoring mobile application (Guardian Connect system) and a third sensor in the arm, which was connected to a glucose sensor recorder (GSR). Subjects were also asked to undergo in-clinic visits of 12-14 h on study days 1, 3, and 7 for frequent blood glucose sample testing using a Yellow Springs Instrument (YSI) reference. RESULTS The overall mean absolute relative difference (MARD ± SD) between abdomen sensor glucose (SG) and YSI reference values was 9.6% ± 9.0% and 9.4% ± 9.8% for the MiniMed 640G insulin pump and Guardian Connect system, respectively; and 8.7% ± 8.0% between arm SG and YSI reference values. The percentage of SG values within 20% agreement of the YSI reference value (for YSI >80 mg/dL) was 90.7% with the MiniMed 640G insulin pump, 91.8% with the Guardian Connect system, and 93.1% for GSR-connected arm sensors. Mean functional sensor life, when calibrating 3-4 times/day, was 145.9 ± 39.3 h for sensors paired with the MiniMed 640G insulin pump, 146.1 ± 41.6 h for sensors paired with the Guardian Connect system, and 147.6 ± 40.4 h for sensors connected to the GSR. Responses to survey questions regarding sensor comfort and ease of use were favorable. CONCLUSIONS The Guardian Sensor 3 glucose sensor, whether located in abdomen or the arm, provided accurate glucose readings when compared with the YSI reference and demonstrated functional life commensurate with the intended 7-day use. ClinicalTrials.gov : NCT02246582.
منابع مشابه
Calibration of Minimally Invasive Continuous Glucose Monitoring Sensors: State-of-The-Art and Current Perspectives
Minimally invasive continuous glucose monitoring (CGM) sensors are wearable medical devices that provide real-time measurement of subcutaneous glucose concentration. This can be of great help in the daily management of diabetes. Most of the commercially available CGM devices have a wire-based sensor, usually placed in the subcutaneous tissue, which measures a "raw" current signal via a glucose-...
متن کاملEvaluation of sensor sites for continuous glucose monitoring in cats with diabetes mellitus.
The continuous glucose monitoring system allows generation of detailed glucose curves via measurement of glucose concentration in interstitial fluid. The conventional site for sensor placement in diabetic cats is the subcutaneous tissue of the lateral chest wall. The aim of this study was to investigate the feasibility and accuracy of sensors placed in the lateral chest wall and in two alternat...
متن کاملContinuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.
BACKGROUND Commercialization of a closed-loop artificial pancreas system that employs continuous subcutaneous insulin infusion and interstitial fluid glucose sensing has been encumbered by state-of-the-art technology. Continuous glucose monitoring (CGM) devices with improved accuracy could significantly advance development efforts. However, the current accuracy of CGM devices might be adequate ...
متن کاملComparing Effects of Continuous Insulin Infusion with or without Subcutaneous Glargine Insulin on Glycemic Control in Diabetic Patients Undergoing Coronary Artery Bypass Graft (CABG)
OBJECTIVE: Hyperglycemia is associated with increased morbidity and mortality in diabetic patients following coronary artery bypass grafting. Tight glycemic control in perioperative period can reduce these events. The goal of this study was to determine whether combination of continuous infusion and subcutaneous glargine as a basal insulin could improve glycemic control. MATERIAL AND METHODS:...
متن کاملReal-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study
Accurate and timely glucose monitoring is essential in intensive care units. Real-time continuous glucose monitoring system (CGMS) has been advocated for many years to improve glycemic management in critically ill patients. In order to determine the effect of calibration time on the accuracy of CGMS, real-time subcutaneous CGMS was used in 18 critically ill patients. CGMS sensor was calibrated ...
متن کامل