Modulation factors calculated with an EPID‐derived MLC fluence model to streamline IMRT/VMAT second checks

نویسندگان

  • Stephen Steciw
  • Satyapal Rathee
  • Brad Warkentin
چکیده

This work outlines the development of a robust method of calculating modulation factors used for the independent verification of MUs for IMRT and VMAT treatments, to replace onerous ion chamber measurements. Two-dimensional fluence maps were calculated for dynamic MLC fields that include MLC interleaf leakage, transmission, and tongue-and-groove effects, as characterized from EPID-acquired images. Monte Carlo-generated dose kernels were then used to calculate doses for a modulated field and that field with the modulation removed at a depth specific to the calculation point in the patient using in-house written software, Mod_Calc. The ratio of these two doses was taken to calculate modulation factors. Comparison between Mod_Calc calculation and ion chamber measurement of modulation factors for 121 IMRT fields yielded excellent agreement, where the mean difference between the two was -0.3% ± 1.2%. This validated use of Mod_Calc clinically. Analysis of 5,271 dynamic fields from clinical use of Mod_Calc gave a mean difference of 0.3% ± 1.0% between Mod_Calc and Eclipse-generated factors. In addition, 99.3% and 96.5% fields pass 5% and 2% criteria, respectively, for agreement between these two predictions. The development and use of Mod_Calc at our clinic has considerably streamlined our QA process for IMRT and RapidArc fields, compared to our previous method based on ion chamber measurements. As a result, it has made it feasible to maintain our established and trusted current in-house method of MU verification, without resorting to commercial software alternatives.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

The aim in this study is to develop a generalized strategy for 3D dose verification of IMRT and VMAT planes using EPID transit images in combination with Monte Carlo (MC) simulations. An EPID-based dosimetric verification procedure was developed to convert EPID-measured transit images into 2D exit photon fluence by de-convoluting with the MC-simulated EPID response kernels. The present scatter ...

متن کامل

Anatomy‐based, patient‐specific VMAT QA using EPID or MLC log files

In this project, we investigated the use of an electronic portal imaging device (EPID), together with the treatment planning system (TPS) and MLC log files, to determine the delivered doses to the patient and evaluate the agreement between the treatment plan and the delivered dose distribution. The QA analysis results are presented for 15 VMAT patients using the EPID measurements, the ScandiDos...

متن کامل

Some considerations concerning volume-modulated arc therapy

In this paper it is formally shown that the dynamic multileaf collimator (MLC) IMRT delivery technique remains valid if the MLC is supported on a 1D moving platform. It is also shown that, in such circumstances, it is always time preferable to deliver overlapping modulating fields as a single swept field rather than as separate fields. The most general formulism is presented and then related to...

متن کامل

WE-E-BRB-07: Direct 3D Fluence Calculation from Machine Beam Parameters for VMAT Delivery Verification.

PURPOSE MLC dynamic log files have been clinically used for quality assurance for years. The logged machine parameters and the derived beam 2D fluence maps can be compared to the ones obtained in the treatment plans in order to evaluate the accuracy and consistency of IMRT beam deliveries. In this study, we propose a computationally efficient method, called Direct 3D Fluence Calculation or D3DF...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2013