The fixion proximal femur nailing system: biomechanical properties of the nail and a cadaveric study.

نویسندگان

  • Ely L Steinberg
  • Nehemia Blumberg
  • Shmuel Dekel
چکیده

The treatment of choice for early mobilization of hip fracture is surgery, which traditionally employs side plates and screws or intramedullary nails. We examined the biomechanical properties of a new proximal femoral nail system. The new expandable Fixion proximal femur nailing (PFN) system, made of stainless-steel alloy, consists of a nail, a peg and an anti-rotation pin. Upon positioning, the nail and peg are expanded to their maximal diameter. The current biomechanical study investigated: nail bending strength and stiffness, fatigue properties and hip peg strength. A cadaveric study that determined the effect of the expandable peg on the femoral head included subsidence testing, pull and torsion testing and intra-osseous pressure (IOP) measurements before and after expansion. Biomechanical properties of the new nail met ASTM F384 guideline requirements. The cadaver study yielded equivalent results for the pullout test between the peg and the hip screw, but found the peg superior in the torsion strength test. IOP during peg insertion and expansion was substantially lower than the threshold pressure that causes avascular necrosis. The biomechanical tests found the new system to be safe and able to provide good abutment of the nail to the bone. We conclude that the Fixion PFN system proved to be an effective proximal femur fracture fixation device.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Expandable intramedullary nail: review of biomechanical studies

This article reviews studies of Fixion expandable nail biomechanical properties in case of long bones fractures osteosynthesis. The purpose of this study was to systematize biomechanical researches dedicated to the expandable nail. There were found seven biomechanical studies published between 2005 and 2008. Among them, six studies are devoted to mechanical experiments, and one is a comment to ...

متن کامل

Treatment of acute tibial shaft fractures with an expandable nailing system: a systematic review of the literature.

Locked Intramedullary Nailing is an established method of treatment for tibial shaft fractures. Locking of tibial nails is however not without its drawbacks. Locking is time consuming, and is associated with a noteworthy complication rate. The Fixion IM nailing system is an expandable nail designed to eliminate some of these drawbacks. We have performed a systematic review of the literature to ...

متن کامل

The use of combined 3.5 LCP unicortical plate and nail fixation in proximal tibia fractures and prevention of valgus and anterior angulation

Introduction: Intramedullary nailing (IMN) of high tibia fracture has some mechanical and biological advantages over the other form of fixation. However, valgus, apex anterior malalignment and anterior displacement of proximal fragment commonly occur after isolated IMN fixation of proximal high tibia fracture. The purpose of this study is to determine the effectiveness of using 3.5-mm locking c...

متن کامل

Elastic Stable Intramedullary Nailing of Femoral Shaft Fracture-Experience in 48 Children

Background: Femoral shaft fractures are an incapacitating pediatric injury accounting for 1.6% of all pediatric bonyinjuries. Management of these fractures is largely directed by age, fracture pattern, associated injuries, built of the childand socioeconomic status of the family. We retrospectively evaluated the use of elastic stable intramedullary nail (ESIN)in surgical management of femoral s...

متن کامل

The Fixion nailing system for stabilising diaphyseal fractures of the humerus: a two-year clinical experience.

Previous studies have shown that the expandable nail system (Fixion) can provide rapid stabilisation of long bone fractures with reduced operative time and low complication rates. Patients with humeral shaft fracture were treated consecutively over a two-year period in our institution with the Fixion nail. Nineteen Fixion nailings were performed in 16 patients over a 2 year period. All fracture...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomechanics

دوره 38 1  شماره 

صفحات  -

تاریخ انتشار 2005