Evaluation of the relation between breast glandular absorbed dose and radiographic quality in mammography
نویسندگان
چکیده
*Corresponding Author: Dr. Mohammad Bagher Tavakoli, Department of Medical Physics and Medical Engineering, Isfahan University of Medical Sciences, Isfahan, Iran. Fax: +98 311 7922412 E-mail: [email protected] Background: Breast is one of the main radiological sensitive organs, hence it is important to evaluate absorbed dose to this organ especially to the glandular parts. It is the aim of this study to measure mean glandular dose and image contrast in terms of different mammographic parameters. Materials and Methods: In this study two mammography machines located at Said-al Shohada (Giotto) and Shahid Behesti (GE) hospitals were used. According to the recommendations of ACR and MQSA, breast phantoms were constructed and used for this study. For dose evaluation TLD dosimetry method was used. The TLD dosimeters were of LiF type and the reader was a Solaro TLD reader. Results: To obtain a constant contrast when increasing kVp from 22 to 24, it was necessary to reduce mAs by 12 percent. The obtained relation between these two parameters is: contrast=0.2829D-0.2427. It was also found that there is a linear relationship between contrast and image quality. The relation between these two parameters is: Image quality = 28.117 Contarst + 20.134. Increasing kVp and hence decreasing mAs results a reduction to the glandular dose, especially in patients with large breast. Increasing kVp from 28 to 30 results in reduction of dose from 6.8mGy to 5mGy. Conclusions: It was found that there has been a linear relationship between contrast and image quality. It was also found that increasing kVp necessitate to reduce mAs for a constant contrast and hence reduction of glandular dose. Iran. J. Radiat. Res., 2008; 6 (2): 7782
منابع مشابه
Evaluation of the relation between breast glandular absorbed dose and radiographic quality in mammography
Background: Breast is one of the main radiological sensitive organs, hence it is important to evaluate absorbed dose to this organ especially to the glandular parts. It is the aim of this study to measure mean glandular dose and image contrast in terms of different mammographic parameters. Materials and Methods: In this study two mammography machines located at Said-al Shohada (Giotto) ...
متن کاملنتایج یک کُد فرترن برای محاسبه دوز غدهای در ماموگرافی با استفاده از پارامترهای Sobol-Wu
Background: Accurate computation of the radiation dose to the breast is essential to mammography. Various the thicknesses of breast, the composition of the breast tissue and other variables affect the optimal breast dose. Furthermore, the glandular fraction, which refers to the composition of the breasts, as partitioned between radiation-sensitive glandular tissue and the adipose tissue, also h...
متن کاملCalculation of Absorbed Glandular Dose using a FORTRAN Program Based on Monte Carlo X-ray Spectra in Mammography
Introduction: Average glandular dose calculation in mammography with Mo-Rh target-filter and dose calculation for different situations is accurate and fast. Material and Methods: In this research, first of all, x-ray spectra of a Mo target bombarded by a 28 keV electron beam with and without a Rh filter were calculated using the MCNP code. Then, we used the Sobol-Wu parameters to write a FORTRA...
متن کاملDetermination of Average Glandular Dose in Mammography for Rh-Rh Mammogram Using Sobol-Wu Parameters in FORTRAN Code
Introduction: To write a computational program to calculate the average glandular dose in Mammography with Rh-Rh target-filter and also use it as an accurate and fast method of dose calculation founder different situations. Materials and Methods: In this research, the Sobol-Wu parameters are used to write a FORTRAN code for glandular dose calculation in Rh-Rh target-filter mammograms. The code ...
متن کاملEvaluation of glandular dose in mammography in the presence of breast cyst using Monte Carlo simulation
Introduction: Average glandular dose (AGD), entrance skin air kerma (ESAK) and normalized glandular dose (DgN) are the main dosimetric quantities in mammography. In this study, DgN is evaluated in the presence of breast cyst, which is a common disease among women and the influence of size, number and location of the cysts on the DgN is investigated. Materials and Meth...
متن کامل