CASE REPORTS A solution to the non-deflating balloon of a suprapubic catheter

نویسندگان

  • T S Huseyin
  • S M Moalypour
چکیده

CASE REPORT A bedbound 62 year old woman attended our department. She had multiple sclerosis and her long term suprapubic catheter was not draining. The catheter balloon would not deflate. The valve mechanism was cut off and a fine guidewire was passed down the balloon’s insufflation passage, but the balloon would still not deflate. The following action was taken. Gentle traction was applied to the catheter, which brought the balloon close to the internal opening of the fistula tract. An intravenous cannula 18G (Green Venflon) was taken and its hub and cap were removed. A 20 ml empty syringe was attached to the sheathed needle. The sheath was advanced over the cannula needle to hide its point. This blunt end was slowly passed down the fistula close to the catheter wall until resistance was met (presumed to be the balloon). The needle tip was advanced through the sheath and the syringe started to automatically fill with the balloon fluid. Once all the fluid was aspirated the catheter was easily removed. The balloon was intact.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A solution to the non-deflating balloon of a suprapubic catheter.

A case report describes the successful aspiration of a non-deflating suprapubic urinary catheter by passing a cannula directly down the fistula tract. It is a relatively safe and simple procedure that can be performed in an accident and emergency department.

متن کامل

A technique for non-deflating balloon catheter removal in female patients

Removing a foley catheter can sometimes be a challenge, especially when it is related to non-deflating balloons which represent most of the cases. In female patients, due to their urethral anatomy, several techniques have been proposed for balloon puncture when other maneuvers have failed. We present a simple technique for non-deflating balloon catheter removal in an old female patient with a p...

متن کامل

Preventable long-term complications of suprapubic cystostomy after spinal cord injury: Root cause analysis in a representative case report

BACKGROUND Although complications related to suprapubic cystostomies are well documented, there is scarcity of literature on safety issues involved in long-term care of suprapubic cystostomy in spinal cord injury patients. CASE PRESENTATION A 23-year-old female patient with tetraplegia underwent suprapubic cystostomy. During the next decade, this patient developed several catheter-related com...

متن کامل

A practical approach to difficult urinary catheterizations.

Routine placement of transurethral catheters can be challenging in some situations, such as urethral strictures, severe phimosis and false passages. Intravaginal retraction of the urethral meatus can complicate Foley placement in postmenopausal females. In men, blind urethral procedures with mechanical or metal sounds without visual guidance or guidewire assistance are now discouraged due to th...

متن کامل

Severe Intraperitoneal Haemorrhage following Suprapubic Catheter Insertion in a Patient Treated with Iloprost

Suprapubic catheter (SPC) insertion is a common urological procedure, performed both in the elective and emergency settings. The authors present an unusual case of severe intraperitoneal bleeding following the insertion of an SPC under direct vision, where the use of prostacyclin analogue may have been a contributing factor.