[Comparison of setup accuracy between ExacTrac X-ray 6 dimensions and cone-beam computed tomography for intracranial and pelvic image-guided radiotherapy].

نویسندگان

  • Tsuyoshi Kudo
  • Kaoru Ono
  • Kengo Furukawa
  • Sachie Fujimoto
  • Yukio Akagi
  • Tadashi Koyama
  • Yutaka Hirokawa
چکیده

The aim of this study was to compare the setup difference measured with ExacTrac X-ray 6D (ETX6D) and cone-beam computed tomography (CBCT) for non-invasive fractionated radiotherapy. Setup data were collected on a Novalis Tx treatment unit for both a head phantom and patients with intracranial tumors and a pelvic phantom and patients with prostate cancer. Initially, setup was done for a phantom using ETX6D. Secondly, a treatment couch was shifted or rotated by each already known value. Thirdly, ETX6D and CBCT scans were obtained. Finally, setup difference was determined: the registrations of ETX6D images with the corresponding digitally reconstructed radiographs using ETX6D fusion, and registrations of CBCT images with the planning CT using online 6D fusion. The setup difference between ETX6D and CBCT was compared. The impact of shifts and rotations on the difference was evaluated. Patients' setup data was similarly analyzed. In phantom experiments, the root mean square (RMS) of difference of the shift and rotation was less than 0.45 mm for translations, and 0.17 degrees for rotations. In intracranial patients' data, the RMS of that was 0.55 mm and 0.44 degree, respectively. In prostate cancer patients' data, the RMS of that was 0.77 mm and 0.79 degree, respectively. In this study, we observed modest setup differences between ETX6D and CBCT. These differences were generally less than 1.00 mm for translations, and 1.00 degrees for rotations, respectively.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Online 6 Degree-of-Freedom Image Registration of Varian TrueBeam Cone-Beam CT and BrainLab ExacTrac X-Ray for Intracranial Radiosurgery

PURPOSE The study was aimed to compare online 6 degree-of-freedom image registrations of TrueBeam cone-beam computed tomography and BrainLab ExacTrac X-ray imaging systems for intracranial radiosurgery. METHODS Phantom and patient studies were performed on a Varian TrueBeam STx linear accelerator (version 2.5), which is integrated with a BrainLab ExacTrac imaging system (version 6.1.1). The p...

متن کامل

Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery

The objective of this study was to evaluate the setup discrepancy between BrainLAB 6 degree-of-freedom (6D) ExacTrac and cone-beam computed tomography (CBCT) used with the imaging guidance system Novalis Tx for intracranial stereotactic radiosurgery. We included 107 consecutive patients for whom white stereotactic head frame masks (R408; Clarity Medical Products, Newark, OH) were used to fix th...

متن کامل

Usability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter

Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...

متن کامل

Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space

Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...

متن کامل

Comparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications

Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased consi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Nihon Hoshasen Gijutsu Gakkai zasshi

دوره 68 12  شماره 

صفحات  -

تاریخ انتشار 2012