Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function.
نویسندگان
چکیده
A discrete representation of the reconstruction process, consistent with the method of data collection, has been used to derive expressions for the noise power spectrum, autocorrelation function and noise equivalent quanta (NEQ) of a computed tomography (CT) image. These parameters have been expressed in terms of basic scanning factors such as tube current, exposure time, slice width and number of detectors. Each of these factors affects the overall magnitude of the noise power spectrum, but the spatial frequency dependence is also determined by the type of reconstruction filter used in the computer algorithm. The noise power spectrum has been calculated for scanners employing either a ramp or Hanning weighted ramp filter. Predictions made from this theoretical analysis have been compared with experimental measurements made on various CT scanners. Measurements were made of the modulation transfer function (MTF) by techniques which permitted us to deduce the contributions of the algorithmic and non-algorithmic components to the overall MTF. NEQ values have been calculated for a number of CT scanners.
منابع مشابه
Impact of Photon Spectra on the Sensitivity of Polymer Gel Dosimetry by X-Ray Computed Tomography
Introduction: The purpose of the current study was to investigate the effect of X-ray spectra on the sensitivity of a polymer gel dosimeter imaged with a conventional computed tomography (CT) scanner. Material and Methods: The whole process of CT imaging of an irradiated polymer gel was simulated by MCNPX Monte Carlo (MC) code. The imaging of polyacrylamide gel was accomplished by means of a co...
متن کاملAssessment of X-Ray Crosstalk in a Computed Tomography Scanner with Small Detector Elements Using Monte Carlo Method
Introduction: Crosstalk is a leakage of X-ray or light produced in a matrix of X-ray detectors or array of photodiodes in one element to other elements affecting on image contrast and spatial resolution. In this study, we assessed X-ray crosstalk in a computed tomography (CT) scanner with small detector elements to estimate the effect of various parameters such as X-ray tube voltage, detector e...
متن کاملShearlet-Based Adaptive Noise Reduction in CT Images
The noise in reconstructed slices of X-ray Computed Tomography (CT) is of unknown distribution, non-stationary, oriented and difficult to distinguish from main structural information. This requires the development of special post-processing methods based on the local statistical evaluation of the noise component. This paper presents an adaptive method of reducing noise in CT images employing th...
متن کاملAssessment the effect of CT localizer parameter on radiation dose and image quality of GE CT spiral images, A phantom study
Introduction: Using computed tomography (CT) is increasing significantly in Iran and worldwide. Reduction of CT's radiation dose is a critical concern. The aim of this study was evaluating the effect of localizer kVp on spiral CT images when the tube current modulation (TCM) is activated. Materials and Methods: A special wedding cake phantom with circular in...
متن کاملInvestigation of dosimetric characteristic of NIPAM polymer gel using x-ray CT
Introduction: Polymer gel dosimeters contain chemical materials sensitive to the radiation which are polymerized by the radiation as a function of absorbed dose. So information of spatial dose distribution can be extracted by imaging from irradiated gel. Among imaging techniques, computed tomography (CT) poses as an attractive method because of practical advantages such as acce...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 29 11 شماره
صفحات -
تاریخ انتشار 1984