Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
نویسندگان
چکیده
OBJECTIVE The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured. RESULTS The plasma glucose range was 2.2-9.4 mmol/L with hematocrit 33-37% and mean HbA(1c) 5.5% ± 0.56 (SD). All meters provided plasma equivalent results except one, which reported whole blood glucose that was adjusted to plasma equivalent values. The absolute glucose difference [meter--plasma glucose] was 0.232 ± 0.69 to 0.725 ± 0.62 mmol/L mean ± SD and bias ranged from 6.1 to 15.8%. Two meters were affected by hematocrit <36% (P < 0.05). CONCLUSIONS Blood glucose meters in current use are not optimally accurate when compared with plasma glucose measurement in DP. Recognition of this deviation is essential to prevent inappropriate treatment of DP.
منابع مشابه
Evolution of analytical performance in portable glucose meters in the last decade.
OBJECTIVE To assess and compare the technical accuracy of portable glucose meters during the last decade. RESEARCH DESIGN AND METHODS One-thousand preprandial (pre) and postprandial (post) capillary whole-blood glucose values measured with meters owned mainly by diabetic patients were compared with a single laboratory method yearly from 1989 to 1999. A total of 21,950 capillary measurements a...
متن کاملDoes random blood glucose sampling outdate testing for glycosuria in the detection of diabetes during pregnancy?
A random blood glucose concentration was determined in 2403 pregnant women attending an antenatal clinic at between 28 and 32 weeks' gestation. The calculated 99% cut off values were 6.1 mmol/l (110 mg/100 ml) within two hours after a meal and 5.6 mmol/l (101 mg/100 ml) more than two hours after a meal. Patients with a blood glucose concentration in excess of these values were referred for a 75...
متن کاملDoes Blood Glucose Monitoring Increase Prior to Clinic Visits in Children With Type 1 Diabetes?
OBJECTIVE To assess the occurrence of white coat adherence in families with children who have type 1 diabetes. RESEARCH DESIGN AND METHODS Blood glucose data were downloaded from meters of 72 children, aged 2-11 years, with type 1 diabetes at four consecutive clinic visits. Generalized estimating equations were used to analyze patterns of blood glucose monitoring (BGM) during the 28 days befo...
متن کاملThe Effect of Probiotic Yogurt on Blood Glucose and cardiovascular Biomarkers in Patients with Type II Diabetes: A Randomized Controlled Trial
Background: Given the high prevalence of type II diabetes and its complications, the evidence regarding the beneficial effects of probiotic yogurt on some cardiovascular biomarkers in diabetic patients is worthy of investigation. Aim: To investigate the effect of probiotic yogurt on blood glucose level and cardiovascular biomarkers in patients with type II diabetes. Method:This randomized, clin...
متن کاملContinued use of an integrated meter with electronic logbook maintains improvements in glycemic control beyond a randomized, controlled trial.
BACKGROUND Blood glucose monitoring is an important component of diabetes self-management for individuals with insulin-treated diabetes. Although patient-maintained logbooks are routinely used, glucose values may be inaccurately recorded or not recorded at all. Electronic logbooks may help overcome such problems. We conducted a randomized, controlled trial (RCT) to compare glycemic control in i...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 34 شماره
صفحات -
تاریخ انتشار 2011