Broad-range direct detection and identification of fungi by use of the PLEX-ID PCR-electrospray ionization mass spectrometry (ESI-MS) system.
نویسندگان
چکیده
The PLEX-ID system is a novel technology that couples PCR amplification and electrospray ionization-mass spectrometry to identify pathogens directly in clinical specimens. The analytical performance of the PLEX-ID Broad Fungal assay was compared with that of traditional culture identification by using 91 characterized fungal culture isolates (64 manufacturer-claimed and 27 nonclaimed organisms) and directly by using 395 respiratory specimens. Discordant results were resolved by D2 large-subunit ribosomal DNA fungal sequencing. Environmental studies were performed to monitor for potential contamination. The PLEX-ID Broad Fungal assay correctly identified 95.6% (87/91) and 81.3% (74/91) of the culture isolates to the genus and species levels, respectively. Of the manufacturer-claimed organisms, 100% (64/64) and 92.2% (59/64) were correctly identified to the genus and species levels, respectively. Direct analysis of respiratory specimens resulted in 67.6% (267/395) and 66.6% (263/395) agreement with culture results to the genus and species levels, respectively, with 16.2% (64/395) of the results discordant with culture and 16.2% (64/395) not detected by the system. The majority (>95%) of the isolates not detected directly by the PLEX-ID system ultimately grew in low quantities in culture (≤ 20 colonies). In 20.3% (35/172) of the respiratory specimens where no growth was observed in culture, the PLEX-ID system identified a fungus, suggesting a potential increase in sensitivity over culture in some instances. The PLEX-ID system provides a rapid method for the detection of a broad array of fungi directly in respiratory specimens and has the potential of impacting turnaround times and patient care by reducing the need to wait for the growth of an organism in culture.
منابع مشابه
Evaluation of the Broad-Range PCR-Electrospray Ionization Mass Spectrometry (PCR/ESI-MS) System and Virus Microarrays for Virus Detection
Advanced nucleic acid-based technologies are powerful research tools for novel virus discovery but need to be standardized for broader applications such as virus detection in biological products and clinical samples. We have used well-characterized retrovirus stocks to evaluate the limit of detection (LOD) for broad-range PCR with electrospray ionization mass spectrometry (PCR/ESI-MS or PLEX-ID...
متن کاملCorrection: Taliaferro, L. et al. Evaluation of the Broad-Range PCR-Electrospray Ionization Mass Spectrometry (PCR/ESI-MS) System and Virus Microarrays for Virus Detection. Viruses 2014, 6, 1876-1896
We have noted some inaccuracies in Table 1 and Table 2 of our article " Evaluation of the Broad-Range PCR-Electrospray Ionization Mass Spectrometry (PCR/ESI-MS) System and Virus Microarrays for Virus Detection " (Viruses 2014, 6, 1876–1896) [1]. In Table 1, the heading " LLMDA " was omitted and needs to be added to correctly align the results obtained using the LLMDA and the ViroChip with their...
متن کاملRapid and High-Throughput Detection of Highly Pathogenic Bacteria by Ibis PLEX-ID Technology
In this manuscript, we describe the identification of highly pathogenic bacteria using an assay coupling biothreat group-specific PCR with electrospray ionization mass spectrometry (PCR/ESI-MS) run on an Ibis PLEX-ID high-throughput platform. The biothreat cluster assay identifies most of the potential bioterrorism-relevant microorganisms including Bacillus anthracis, Francisella tularensis, Ye...
متن کاملIdentification of Clinically Relevant Fungi and Prototheca Species by rRNA Gene Sequencing and Multilocus PCR Coupled with Electrospray Ionization Mass Spectrometry
BACKGROUND Multilocus PCR coupled with electrospray ionization mass spectrometry (PCR/ESI-MS) is a new strategy for pathogen identification, but information about its application in fungal identification remains sparse. METHODS One-hundred and twelve strains and isolates of clinically important fungi and Prototheca species were subjected to both rRNA gene sequencing and PCR/ESI-MS. Three regi...
متن کاملQuantification of Melittin in Iranian Honey Bee (Apis mellifera meda) Venom by Liquid Chromatography-electrospray Ionization-ion Trap Tandem Mass Spectrometry (LC-ESI-IT-MS/MS)
The current research aimed to quantify melittin (MEL) in Iranian honey bee (Apis mellifera meda) venom. To this end, a liquid chromatography-electrospray ionization-ion trap tandem mass spectrometry (LC-ESI-IT-MS/MS) approach was employed. Melittin is the main toxic peptide of honey bee venom with various biological and pharmacological activities. It was extracted with...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 51 6 شماره
صفحات -
تاریخ انتشار 2013