Evaluation of semiempirical VMAT dose reconstruction on a patient dataset based on biplanar diode array measurements
نویسندگان
چکیده
We report the results of a preclinical evaluation of recently introduced commercial tools for 3D patient IMRT/VMAT dose reconstruction, the Delta4 Anatomy calculation algorithm. Based on the same initial measurement, volumetric dose can be reconstructed in two ways. Three-dimensional dose on the Delta4 phantom can be obtained by renormalizing the planned dose distribution by the measurement values (D4 Interpolation). Alternatively, incident fluence can be approximated from the phantom measurement and used for volumetric dose calculation on an arbitrary (patient) dataset with a pencil beam algorithm (Delta4 PB). The primary basis for comparison was 3D dose obtained by previously validated measurement-guided planned dose perturbation method (ACPDP), based on the ArcCHECK dosimeter with 3DVH software. For five clinical VMAT plans, D4 Interpolation agreed well with ACPDP on a homogeneous cylindrical phantom according to gamma analysis with local dose-error normalization. The average agreement rates were 98.2% ± 1.3% (1 SD), (range 97.0%-100%) and 92.8% ± 3.9% (89.5%-99.2%), for the 3%/3 mm and 2%/2 mm criteria, respectively. On a similar geometric phantom, D4 PB demonstrated substantially lower agreement rates with ACPDP: 88.6% ± 6.8% (81.2%-96.1%) and 72.4% ± 8.4% (62.1%-81.1%), for 3%/3 mm and 2%/2 mm, respectively. The average agreement rates on the heterogeneous patients' CT datasets are lower yet: 81.2% ± 8.6% (70.4%-90.4%) and 64.6% ± 8.4% (56.5%-74.7%), respectively, for the same two criteria sets. For both threshold combinations, matched analysis of variance (ANOVA) multiple comparisons showed statistically significant differences in mean agreement rates (p < 0.05) for D4 Interpolation versus ACPDP on one hand, and D4 PB versus ACPDP on either cylindrical or patient dataset on the other hand. Based on the favorable D4 Interpolation results for VMAT plans, the resolution of the reconstruction method rather than hardware design is likely to be responsible for D4 PB limitations.
منابع مشابه
Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT
Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...
متن کاملValidation of a GPU‐Based 3D dose calculator for modulated beams
A superposition/convolution GPU-accelerated dose computation algorithm (the Calculator) has been recently incorporated into commercial software. The algorithm requires validation prior to clinical use. Three photon energies were examined: conventional 6 MV and 15 MV, and 10 MV flattening filter free (10 MVFFF). For a set of IMRT and VMAT plans based on four of the five AAPM Practice Guideline 5...
متن کاملMeasurement‐guided volumetric dose reconstruction for helical tomotherapy
It was previously demonstrated that dose delivered by a conventional linear accelerator using IMRT or VMAT can be reconstructed - on patient or phantom datasets - using helical diode array measurements and a technique called planned dose perturbation (PDP). This allows meaningful and intuitive analysis of the agreement between the planned and delivered dose, including direct comparison of the d...
متن کاملInitial dosimetric evaluation of SmartArc – a novel VMAT treatment planning module implemented in a multi‐vendor delivery chain
We performed an initial dosimetric evaluation of SmartArc - a novel VMAT planning module for Philips Pinnacle treatment planning system. A test suite of structure sets and dose objectives provided by the AAPM for multi-institutional comparison of IMRT dosimetry was used. A total of fifty plans were successfully delivered. The effect of control point spacing on dosimetric accuracy was investigat...
متن کاملCommissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA
We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response ...