The impact of arm position and pulse pressure on the validation of a wrist-cuff blood pressure measurement device in a high risk population
نویسندگان
چکیده
Despite the increasing popularity of blood pressure (BP) wrist monitors for self-BP measurement at home, device validation and the effect of arm position remains an issue. This study focused on the validation of the Omron HEM-609 wrist BP device, including an evaluation of the impact of arm position and pulse pressure on BP measurement validation. Fifty patients at high risk for cardiovascular disease were selected (age 65 +/- 10 years). Each patient had two measurements with a mercury sphygmomanometer and three measurements with the wrist BP device (wrist at the heart level while the horizontal arm supported [HORIZONTAL], hand supported on the opposite shoulder [SHOULDER], and elbow placed on a desk [DESK]), in random order. The achieved systolic BP (SBP) and diastolic BP (DBP) wrist-cuff readings were compared to the mercury device and the frequencies of the readings within 5, 10, and 15 mmHg of the gold standard were computed and compared with the British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocols. The results showed while SBP readings with HORIZONTAL and SHOULDER positions were significantly different from the mercury device (mean difference = 7.1 and 13.3 mmHg, respectively; P < 0.05), the DESK position created the closest reading to mercury (mean difference = 3.8, P > 0.1). Approximately 71% of SBP readings with the DESK position were within +/-10 mmHg, whereas it was 62.5% and 34% for HORIZONTAL and SHOULDER positions, respectively. Wrist DBP attained category D with BHS criteria with all three arm positions. Bland-Altman plots illustrated that the wrist monitor systematically underestimated SBP and DBP values. However a reading adjustment of 5 and 10 mmHg for SBP and DBP (DESK position) resulted in improvement with 75% and 77% of the readings being within 10 mmHg (grade B), respectively. AAMI criteria were not fulfilled due to heterogeneity. The findings also showed that the mismatch between the mercury and wrist-cuff systolic BP readings was directly associated with pulse pressure. In conclusion the DESK position produces the most accurate readings when compared to the mercury device. Although wrist BP measurement may underestimate BP measured compared to a mercury device, an adjustment by 5 and 10 mmHg for SBP and DBP, respectively, creates a valid result with the DESK position. Nevertheless, considering the observed variations and the possible impact of arterial stiffness, individual clinical validation is recommended.
منابع مشابه
تخمین بدون کاف فشارخون مبتنی بر ویژگیهای زمانی سیگنال نبض
Blood pressure is one of the vital signs. Specially, it is crucial for some cases such as hypertension patients and it should be monitored continuously in ICU/CCU. It must be noted that current systems to measure blood pressure, often require trained operators. As an example, in post-hospital cares, blood pressure control is difficult except with the presence of a nurse or use of a device that ...
متن کاملContinuous and Cuffless Blood Pressure Monitoring Based on ECG and SpO2 Signals By Using Microsoft Visual C Sharp
Background: One of the main problems especially in operating room and monitoring devices is measurement of Blood Pressure (BP) by sphygmomanometer cuff.Objective: In this study we designed a new method to measure BP changes continuously for detecting information between cuff inflation times by using vital signals in monitoring devices. This will be achieved by extraction of the time difference ...
متن کاملتأثیر وضعیتهای مختلف قرارگیری اندامهای بیماران مبتلا به پرفشاری خون بر میزان فشارخون آنان
Introduction: In hypertensive patients M position (sitting, right arm at the level of heart and both plantar surface of foots on the floor) is the standard position for measurment of blood pressure since it reveals the “true” pressure. Materials and Methods: It is a quaziexperimental study that compares effect of different positions of limbs on blood pressure of hypertensive patients. The sampl...
متن کاملComparison of blood pressure measurements on the upper and lower extremities during spinal anesthesia
Introduction: During spinal anesthesia, measurement of blood pressure is often obtained by an indirect method using an inflatable cuff on the upper arm. The relationship between the blood pressure measurements at the two sites in patients undergoing spinal anesthesia is unclear. This study was designed to evaluate the difference observed in systolic blood pressure (SBP) taken from the upp...
متن کاملInaccuracy of wrist-cuff oscillometric blood pressure devices: an arm position artefact?
BACKGROUND Despite the increasing popularity of wrist-cuff blood pressure (BP) devices, their accuracy has not been established and international guidelines do not support their use. Because arm position influences BP measurement, it is possible that conflicting reports on wrist-cuff device accuracy reflects diverse arm positions. METHOD This study compared BP measured by two oscillometric de...