Quadrature Lowpass Birdcage Coil for Low Field Open Mri Scanner
نویسندگان
چکیده
The main goal of this research was to demonstrate the utility of using a quadrature birdcage coil in lowpass version for magnetic resonance imaging (MRI) in open scanner at very low static field. When used in horizontal bore MRI systems, the birdcage coil is able to produce circular polarized field using quadrature excitation, thus reaching SNR improvement of 2 as maximum value. The birdcage coil in high-pass version can be employed even in a vertical bore MRI scanner, by combining the sinusoidal and the end-ring resonant modes. In this paper we investigated the lowpass birdcage coil configuration to operate with a vertical bore MRI system, producing two 1 B fields components perpendicular to the vertical 0 B field. Its implementation is convenient because it requires one half capacitors with respect to the high-pass design. Moreover, at low field the high near-electric field losses, that represent the main disadvantage of the low pass design, can be neglected. Experiments performed on a quadrature birdcage coil prototype mounted on an open MRI scanner, showed that both SNR and homogeneity degree increase with respect to the linear birdcage.
منابع مشابه
Design of a Coil System for Hyperpolarized 3He Human Lung Imaging
Introduction: Clinic hyperpolarized He gas human lung imaging has been widely conducted in recent years. Multi-channel coil designed for human lung parallel imaging has been reported [1] to accelerate the imaging speed. Large asymmetric birdcage coil were also developed to replace the single loop coil for better B1 homogeneity [2]. This work presents a design of a coil system for hyperpolarized...
متن کاملA Nested Dual Frequency Birdcage/Stripline Coil for Sodium/Proton Brain Imaging at 7T
Introduction: Dual tuned sodium/proton coils are valuable for multi-nuclear magnetic resonance imaging and spectroscopy. Due to low sodium content in the body and its low gyromagnetic ratio, the critical design aim for a sodium/proton coil is to maximize sodium sensitivity and homogeneity while maintaining adequate proton sensitivity. This has previously been achieved using a dual-tuned birdcag...
متن کاملThe intrinsic magnetic field symmetries of the spiral birdcage coil
INTRODUCTION Transmit array spatial encoding (TRASE) is a novel MRI technique that utilizes B1-field phase gradients as the means for traversing k-space [1-3]. This method allows MRI to be performed without the usual application of B0-field amplitude gradients, thereby possibly eliminating the equipment, complexity, power consumption and acoustic noise associated with traditional static-field g...
متن کاملDesign and Simulation of a Birdcage Coil using CST Studio Suite for Application at 7T
This work describes the study of coil for Magnetic Resonance Imaging (MRI) application. Concretely, the principal objective is the design of a birdcage coil (RF coil) to use in a 7 Tesla scanner. More strength field has a better SNR and increased chemical shift effects, improving spectral fat suppression and spectroscopy. Moreover, a better SNR increases the spatial resolution or reduces the im...
متن کاملComputational analysis of the high pass birdcage resonator: finite difference time domain simulations for high-field MRI.
In this work, a finite difference time domain (FDTD) algorithm is validated at 1.5 tesla using the standard GE Signa quadrature head coil and a muscle phantom. The electrical characteristics of the birdcage head coil are then calculated for the linear and quadrature cases. Unlike previous computational analysis which assume idealized currents on the end rings and the struts of the resonator, a ...
متن کامل