SU-E-J-170: Clinical Evaluation of Positioning Accuracy of Two Immobilization Devices for Stereotactic Body Radiotherapy Using Cone Beam CT.

نویسندگان

  • M Cao
  • N Cao
  • H Cardenes
  • A Fakiris
  • C Desrosiers
  • I Das
چکیده

PURPOSE To evaluate the patient positioning accuracy and reproducibility of two commercially available immobilization systems for Stereotactic Body Radiotherapy (SBRT) treatment. METHODS Forty one patients with lung (n=21) or liver (n=20) malignancies were assigned to one of the two immobilization devices: Elekta stereotactic body frame (SBF) with built-in stereotactic coordinate system and Civco modular indexing based frame (MIF) without stereotactic reference. All patients underwent the same simulation and planning procedure followed by cone beam CT (CBCT) guided treatment setup. A total of 151 CBCT images were analyzed. The systematic and random isocenter setup errors of the two systems were calculated and compared based on the daily setup corrections under CBCT guidance. RESULTS There was not statistically significant difference between the two systems in terms of systematic setup errors in all three translational directions, for both lung and liver patients. The random errors for the lung patients under SBF setup were 1.8mm, 2.0mm and 2.9mm for the vertical, longitudinal and lateral directions, respectively compared to 3.6mm, 4.1mm, and 4.2mm for MIF. A similar trend was also observed for liver patients. The random errors of liver MIF setup reached 3.5mm, 6.1mm and 5.7mm for the vertical, longitudinal and lateral directions, respectively, with relatively smaller errors 1.7mm, 3.4mm and 2.6mm with SBF setup. Repeated CBCTs occurred for MIF system in 42.4% and 40.7% of the lung and liver treatment to verify couch corrections based on the institutional tolerance, resulting in prolonged setup time. Only 25% and 13.6% of the lung and liver treatment with SBF required with repeated CBCT. CONCLUSIONS Without stereotactic coordinate reference, the body frame system tended to have larger random setup errors and patient positioning accuracy inevitably relies on the volumetric imaging guidance. Patient comfort and reproducibility should be clearly considered for selecting a system.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System

Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: ...

متن کامل

To frame or not to frame? Cone‐beam CT‐based analysis of head immobilization devices specific to linac‐based stereotactic radiosurgery and radiotherapy

PURPOSE Noninvasive frameless systems are increasingly being utilized for head immobilization in stereotactic radiosurgery (SRS). Knowing the head positioning reproducibility of frameless systems and their respective ability to limit intrafractional head motion is important in order to safely perform SRS. The purpose of this study was to evaluate and compare the intrafractional head motion of a...

متن کامل

Evaluation of initial setup errors of two immobilization devices for lung stereotactic body radiation therapy (SBRT)

The aim of this study was to investigate the accuracy and efficacy of two commonly used commercial immobilization systems for stereotactic body radiation therapy (SBRT) in lung cancer. This retrospective study assessed the efficacy and setup accuracy of two immobilization systems: the Elekta Body Frame (EBF) and the Civco Body Pro-Lok (CBP) in 80 patients evenly divided for each system. A cone ...

متن کامل

Evaluation of the tumor movement and the reproducibility of two different immobilization setups for image-guided stereotactic body radiotherapy of liver tumors

BACKGROUND The purpose of this study is to evaluate the tumor movement and accuracy of patient immobilization in stereotactic body radiotherapy of liver tumors with low pressure foil or abdominal compression. METHODS Fifty-four liver tumors treated with stereotactic body radiotherapy were included in this study. Forty patients were immobilized by a vacuum couch with low pressure foil, 14 pati...

متن کامل

Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space

Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part8  شماره 

صفحات  -

تاریخ انتشار 2012