Effects of ventilator settings, nebulizer and exhalation port position on albuterol delivery during non-invasive ventilation: an in-vitro study

نویسندگان

  • Yuda Sutherasan
  • Lorenzo Ball
  • Pasquale Raimondo
  • Valentina Caratto
  • Elisa Sanguineti
  • Federico Costantino
  • Maurizio Ferretti
  • Robert M Kacmarek
  • Paolo Pelosi
چکیده

BACKGROUND Few studies have investigated the factors affecting aerosol delivery during non-invasive ventilation (NIV). Our aim was to investigate, using a bench-top model, the effect of different ventilator settings and positions of the exhalation port and nebulizer on the amount of albuterol delivered to a lung simulator. METHODS A lung model simulating spontaneous breathing was connected to a single-limb NIV ventilator, set in bi-level positive airway pressure (BIPAP) with inspiratory/expiratory pressures of 10/5, 15/10, 15/5, and 20/10 cmH2O, or continuous positive airway pressure (CPAP) of 5 and 10 cmH2O. Three delivery circuits were tested: a vented mask with the nebulizer directly connected to the mask, and an unvented mask with a leak port placed before and after the nebulizer. Albuterol was collected on a filter placed after the mask and then the delivered amount was measured with infrared spectrophotometry. RESULTS Albuterol delivery during NIV varied between 6.7 ± 0.4% to 37.0 ± 4.3% of the nominal dose. The amount delivered in CPAP and BIPAP modes was similar (22.1 ± 10.1 vs. 24.0 ± 10.0%, p = 0.070). CPAP level did not affect delivery (p = 0.056); in BIPAP with 15/5 cmH2O pressure the delivery was higher compared to 10/5 cmH2O (p = 0.033) and 20/10 cmH2O (p = 0.014). Leak port position had a major effect on delivery in both CPAP and BIPAP, the best performances were obtained with the unvented mask, and the nebulizer placed between the leak port and the mask (p < 0.001). CONCLUSIONS In this model, albuterol delivery was marginally affected by ventilatory settings in NIV, while position of the leak port had a major effect. Nebulizers should be placed between an unvented mask and the leak port in order to maximize aerosol delivery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Albuterol delivery during noninvasive ventilation.

HYPOTHESIS Albuterol delivered during noninvasive positive-pressure ventilation is affected by use of a nebulizer or metered-dose inhaler (MDI) and whether the leak port is in the hose or the mask. METHODS A lung model that simulated spontaneous breathing at 20 breaths/min was used. A bi-level positive-airway-pressure ventilator (Respironics S/T30) was set for an inspiratory positive airway p...

متن کامل

Bronchodilator delivery during simulated pediatric noninvasive ventilation.

BACKGROUND Noninvasive ventilation (NIV) is usually applied using bi-level positive airway pressure devices, and many of these devices use a single-limb patient circuit with an integrated leak port to purge the circuit of exhaled carbon dioxide. Sometimes bronchodilator therapy is indicated in pediatric patients, but there have been no studies of the optimal nebulizer position, with respect to ...

متن کامل

Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation.

BACKGROUND The effectiveness of aerosol drug delivery during mechanical ventilation is influenced by the patient, ventilator, and nebulizer variables. The impact of nebulizer type, position on the ventilator circuit, and bias flow on aerosol drug delivery has not been established for different age populations. OBJECTIVE To determine the influence of nebulizer position and bias flow with a jet...

متن کامل

Albuterol delivery via intrapulmonary percussive ventilator and jet nebulizer in a pediatric ventilator model.

BACKGROUND In-line administration of bronchodilators is widely used in pediatric patients receiving mechanical ventilation. We compared the amount of albuterol captured at the end of the endotracheal tube (ETT) with an intrapulmonary percussive ventilator (IPV) versus a Salter 8900 jet nebulizer placed in-line in a pediatric ventilator model, under various operating conditions. We hypothesized ...

متن کامل

Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation.

BACKGROUND The position of the jet or ultrasonic nebulizer in the ventilator circuit impacts drug delivery during mechanical ventilation, but has not been extensively explored, and no study has examined all of the commonly used nebulizers. METHODS Drug delivery from jet, vibrating-mesh, and ultrasonic nebulizers and pressurized metered-dose inhaler (pMDI) with spacer was compared in a model o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2015