Assessment of cardiac output changes using a modified FloTrac/Vigileo™ algorithm in cardiac surgery patients
نویسندگان
چکیده
INTRODUCTION The FloTrac/Vigileo (Edwards Lifesciences, Irvine, CA, USA) allows pulse pressure-derived cardiac output measurement without external calibration. Software modifications were performed in order to eliminate initially observed deficits. The aim of this study was to assess changes in cardiac output determined by the FloTrac/Vigileo system (FCO) with an initially released (FCOA) and a modified (FCOB) software version, as well as changes in cardiac output from the PiCCOplus system (PCO; Pulsion Medical Systems, Munich, Germany). Both devices were compared with cardiac output measured by intermittent thermodilution (ICO). METHODS Cardiac output measurements were performed in patients after elective cardiac surgery. Two sets of data (A and B) were obtained using FCOA and FCOB in 50 patients. After calibration of the PiCCOplus system, triplicate FCO and PCO values were recorded and ICO was determined in the supine position and cardiac output changes due to body positioning were recorded 15 minutes later (30 degrees head-up, 30 degrees head-down, supine). Student's t test, analysis of variance and Bland-Altman analysis were calculated. RESULTS Significant changes of FCO, PCO and ICO induced by body positioning were observed in both data sets. For set A, DeltaFCOA was significantly larger than DeltaICO induced by positioning the head down. For set B, there were no significant differences between DeltaFCOB and DeltaICO. For set A, increased limits of agreement were found for FCOA-ICO when compared with PCO-ICO. For set B, mean bias and limits of agreement were comparable for FCOB-ICO and PCO-ICO. CONCLUSIONS The modification of the FloTrac/Vigileo system resulted in an improved performance in order to reliably assess cardiac output and track the related changes in patients after cardiac surgery.
منابع مشابه
Comparison between Flotrac-Vigileo and Bioreactance, a totally noninvasive method for cardiac output monitoring
INTRODUCTION This study was designed to compare the clinical acceptability of two cardiac output (CO) monitoring systems: a pulse wave contour-based system (FloTrac-Vigileo) and a bioreactance-based system (NICOM), using continuous thermodilution (PAC-CCO) as a reference method. METHODS Consecutive patients, requiring PAC-CCO monitoring following cardiac surgery, were also monitored by the tw...
متن کاملCLINICAL SCIENCE Clinical evaluation of the flotrac/vigileo system for continuous cardiac output monitoring in patients undergoing regional anesthesia for elective cesarean section: a pilot study
BACKGROUND: Spinal anesthesia for cesarean delivery may cause severe maternal hypotension and a decrease in cardiac output. Compared to assessment of cardiac output via a pulmonary artery catheter, the FloTrac/Vigileo system may offer a less invasive technique. The aim of this study was to evaluate cardiac output and other hemodynamic measurements made using the FloTrac/Vigileo system in patien...
متن کاملSelf-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability?
The present study was performed to test a new software version of the FloTrac/Vigileo using head-up-head-down tilting in post-cardiac surgery patients. Impressive improvements in Bland and Altman limits of agreement from 37.5% to 21.6% were recorded. The results, however could be attributed to a failure to produce a wide enough range of test circulatory conditions. A more rigorous test of perfo...
متن کاملClinical evaluation of the flotrac/vigileo™ system for continuous cardiac output monitoring in patients undergoing regional anesthesia for elective cesarean section: a pilot study
BACKGROUND Spinal anesthesia for cesarean delivery may cause severe maternal hypotension and a decrease in cardiac output. Compared to assessment of cardiac output via a pulmonary artery catheter, the FloTrac/Vigileo system may offer a less invasive technique. The aim of this study was to evaluate cardiac output and other hemodynamic measurements made using the FloTrac/Vigileo system in patient...
متن کاملPitfalls in haemodynamic monitoring based on the arterial pressure waveform
The accuracy of the arterial pressure-based cardiac output FloTrac-Vigileo system remains unacceptably low during haemodynamic instability. Data show that the measurement of cardiac output (CO) is strongly influenced by changes in factors that affect arterial blood pressure (ABP) - for example, vascular tone and compliance and the arterial site - independently of true changes in CO. Although in...