A novel test method for evaluation of the abrasive wear behaviour of total hip stems at the interface between implant surface and bone cement.
نویسندگان
چکیده
After total hip replacement, some cemented titanium stems show above-average early loosening rates. Increased release of wear particles and resulting reaction of the peri-prosthetic tissue were considered responsible. The objective was to develop a test method for analysing the abrasive wear behaviour of cemented stems and for generating wear particles at the interface with the bone cement. By means of the novel test device, cemented hip stems with different designs, surface topographies and material compositions using various bone cements could be investigated. Before testing, the cemented stems were disconnected from the cement mantle to simulate the situation of stem loosening (debonding). Subsequently, constant radial contact pressures were applied on to the stem surface by a force-controlled hydraulic cylinder. Oscillating micromotions of the stem (+/- 250 microm; 3 x 10(6)cycles; 5 Hz) were carried out at the cement interface initiating the wear process. The usability of the method was demonstrated by testing geometrically identical Ti-6A1-7Nb and Co-28Cr-6Mo hip stems (n= 12) with definite rough and smooth surfaces, combined with commercially available bone cement containing zirconium oxide particles. Under identical frictional conditions with the rough shot-blasted stems, clearly more wear particles were generated than with the smooth stems, whereas the material composition of the hip stems had less impact on the wear behaviour.
منابع مشابه
110189 NV-MOJ May 08 01-07 G5H.qxd
Total hip replacement has consistently been shown to be one of the most reliable and successful surgical procedures in orthopaedic surgery . However time, the longevity of total hip replacement implants may be limited by osteolysis and aseptic loosening regardless of the method of fixation . Osteolysis manifests itself by causing progressive periprosthetic bone resorption secondary to the biolo...
متن کاملCement-Implant Interface Fracture Failure by Crack Initiation Due to Interface Cavity Stress Concentration
Nowadays total joint replacements are widely used in the world, so in average 800,000 joint surgeries are done yearly only in Europe and North America. However implant loosening is and remains as the major issue of all implant failures and therefore causes revision surgery procedures. Studies and experiments have identified poor fixation of implants most likely is the main cause of long term im...
متن کاملCell viability, collagen synthesis and cytokine expression in human osteoblasts following incubation with generated wear particles using different bone cements.
In total hip arthroplasty, wear particles generated at articulating surfaces and interfaces between bone, cement and implants have a negative impact on osteoblasts, leading to osteolysis and implant loosening. The aim of this experimental study was to determine the effects of particulate wear debris generated at the interface between straight stainless steel hip stems (Exeter(®)) and three diff...
متن کاملThe potential role of human osteoblasts for periprosthetic osteolysis following exposure to wear particles.
Aseptic loosening in total hip replacement is mainly caused by wear particles inducing inflammation and osteolysis. Wear can be a consequence of micromotions at the interface between implant and bone cement. Due to complex cellular interactions, different mediators (e.g. cytokines, proteinases) are released, which can promote osteolytic processes in the periprosthetic tissue followed by looseni...
متن کاملDry and Wet Wear Characteristic of TiO2 Thin Film Prepared by Magnetic Sputtering in Ringer Solution
In this research, a thin film of TiO2 was applied on AZ91D using the method of magnetic sputtering. Microstructure investigations were done using field emission scanning electron microscope (FESEM) and X-ray diffraction (XRD). Wear resistance for the coating was investigated using the pin on the disk in the form of dry and in the Ringer's solution. After this test, the worn surface of the sampl...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
دوره 218 4 شماره
صفحات -
تاریخ انتشار 2004