Experimental Comparison of Knife-Edge and Multi-Parallel Slit Collimators for Prompt Gamma Imaging of Proton Pencil Beams
نویسندگان
چکیده
More and more camera concepts are being investigated to try and seize the opportunity of instantaneous range verification of proton therapy treatments offered by prompt gammas emitted along the proton tracks. Focusing on one-dimensional imaging with a passive collimator, the present study experimentally compared in combination with the first, clinically compatible, dedicated camera device the performances of instances of the two main options: a knife-edge slit (KES) and a multi-parallel slit (MPS) design. These two options were experimentally assessed in this specific context as they were previously demonstrated through analytical and numerical studies to allow similar performances in terms of Bragg peak retrieval precision and spatial resolution in a general context. Both collimators were prototyped according to the conclusions of Monte Carlo optimization studies under constraints of equal weight (40 mm tungsten alloy equivalent thickness) and of the specificities of the camera device under consideration (in particular 4 mm segmentation along beam axis and no time-of-flight discrimination, both of which less favorable to the MPS performance than to the KES one). Acquisitions of proton pencil beams of 100, 160, and 230 MeV in a PMMA target revealed that, in order to reach a given level of statistical precision on Bragg peak depth retrieval, the KES collimator requires only half the dose the present MPS collimator needs, making the KES collimator a preferred option for a compact camera device aimed at imaging only the Bragg peak position. On the other hand, the present MPS collimator proves more effective at retrieving the entrance of the beam in the target in the context of an extended camera device aimed at imaging the whole proton track within the patient.
منابع مشابه
Time-resolved imaging of prompt-gamma rays for proton range verification using a knife-edge slit camera based on digital photon counters.
Proton range monitoring may facilitate online adaptive proton therapy and improve treatment outcomes. Imaging of proton-induced prompt gamma (PG) rays using a knife-edge slit collimator is currently under investigation as a potential tool for real-time proton range monitoring. A major challenge in collimated PG imaging is the suppression of neutron-induced background counts. In this work, we pr...
متن کاملDerivation and validation of a sensitivity formula for knife-edge slit gamma camera: A theoretical and Monte Carlo simulation study
Introduction: Gamma cameras are proposed for online range verification and treatment monitoring in proton therapy. An Analytical formula was derived and validated for sensitivity of a slit collimator based on the photon fluence concept. Methods: Fluence formulation was generalized for photons distribution function and solved for high-energy point sources. The...
متن کاملReal-time prompt γ monitoring in spot-scanning proton therapy using imaging through a knife-edge-shaped slit.
In this paper we report on Monte Carlo simulations to investigate real-time monitoring of the track depth profile in particle therapy by measuring prompt gamma ray emissions: a high sensitivity imaging system employing a knife-edge-shaped slit combined with a position-sensitive gamma detector was evaluated. Calculations to test this new concept were performed for a head-sized software phantom. ...
متن کاملGamma Knife Simulation Using the MCNP4C Code and the Zubal Phantom and Comparison with Experimental Data
Introduction: Gamma Knife is an instrument specially designed for treating brain disorders. In Gamma Knife, there are 201 narrow beams of cobalt-60 sources that intersect at an isocenter point to treat brain tumors. The tumor is placed at the isocenter and is treated by the emitted gamma rays. Therefore, there is a high dose at this point and a low dose is delivered to the normal tissue surroun...
متن کاملImaging of prompt gamma rays emitted during delivery of clinical proton beams with a Compton camera: feasibility studies for range verification.
The purpose of this paper is to evaluate the ability of a prototype Compton camera (CC) to measure prompt gamma rays (PG) emitted during delivery of clinical proton pencil beams for prompt gamma imaging (PGI) as a means of providing in vivo verification of the delivered proton radiotherapy beams. A water phantom was irradiated with clinical 114 MeV and 150 MeV proton pencil beams. Up to 500 cGy...
متن کامل