Taper Failure After Large-Diameter Metal-on-Metal Total Hip Arthroplasty.
نویسندگان
چکیده
The use of metal-on-metal bearing surfaces for total hip arthroplasty increased dramatically in the past decade for a variety of reasons, including the advantage of using larger-diameter femoral heads to reduce instability. However, recent research suggesting negative outcomes and high failure rates associated with some of these implants has decreased their use. Further, the use of larger-diameter femoral heads on relatively small femoral neck tapers has caused concern about localized corrosion at this junction, particularly with varus femoral alignment and longer femoral neck and head lengths. Although the advent of modular components offers the surgeon greater intraoperative flexibility, this modularity may prove to be a weak link when coupled with large-diameter femoral heads. This report describes a patient with a history of bilateral metal-on-metal total hip arthroplasty complicated by failure of the right hip as a result of fretting and mechanically assisted crevice corrosion. Notching of the femoral neck and head dissociation occurred 10 years after the primary procedure. The authors describe the patient's presenting symptoms, the possible etiology of the catastrophic failure, and the method of treatment and provide a brief review of metal-on-metal implants that may shed some light on the complications in this case. [Orthopedics.2016; 39(5):e984-e987.].
منابع مشابه
Taper junction failure in large-diameter metal-on-metal bearings
OBJECTIVES An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer. METHODS Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular...
متن کاملDamage Patterns at the Head-Stem Taper Junction Helps Understand the Mechanisms of Material Loss.
BACKGROUND Material loss at the taper junction of metal-on-metal total hip arthroplasties has been implicated in their early failure. The mechanisms of material loss are not fully understood; analysis of the patterns of damage at the taper can help us better understand why material loss occurs at this junction. METHODS We mapped the patterns of material loss in a series of 155 metal-on-metal ...
متن کاملClinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty
OBJECTIVES Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of materi...
متن کاملFemoral diameter and stem type are independent risk factors for ARMD in the Large-headed ASR THR group
BACKGROUND Adverse soft-tissue reaction to metal debris (ARMD) continues to be major source of concern in metal-on-metal (MoM) hip replacements. In our earlier study we were able to establish several risk factors for ARMD in patients who had received a small-diameter (<50 mm) Articular Surface Replacement (ASR, DePuy, Warsaw, IN, USA). The aims of the present study were to analyze whether these...
متن کاملCurrent Insights Regarding Metal-On-Metal Bearings for Hip Arthroplasty "2279
Modern small diameter metal-on-metal (MoM) bearings for total hip arthroplasty (THA) have been developed in the nineteen-eighties to address the problem of polyethylene wear related osteolysis. Subsequently large diameter MoM hip resurfacings (HRA) were designed for young and active patients to preserve bone and avoid dislocation. Large diameter MoM THA were originally meant as an easy femoral ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Orthopedics
دوره 39 5 شماره
صفحات -
تاریخ انتشار 2016