Analytical and clinical performance of improved Abbott IMx CA 125 assay: comparison with Abbott CA 125 RIA.
نویسندگان
چکیده
We compared the improved Abbott IMx cancer antigen (CA) 125 assay (cat. no. 7A89) with the Abbott CA 125 RIA. Serum specimens were from healthy perimenopausal women (n = 124) and from patients with benign gynecologic and nongynecologic diseases (n = 124), ovarian carcinoma (n = 104), or other malignancies (n = 193). The IMx assay detected as little as 0.193 kAU/L CA 125 (AU = arbitrary Abbott unit), demonstrated up to 29% overestimation upon serum dilution, low within-assay (2.7-5.6%) and between-assay (4.8-8.2%) CVs, and no high-dose hook effect < or = 46,000 kAU/L nor influence from human anti-mouse antibodies in serum of women injected with OC 125 F(ab')2. Values by IMx were 20% lower than by RIA for healthy perimenopausal women (n = 100; IMx = 0.80 RIA - 2.5 kAU/L), and at least 50% higher for those with benign or malignant ovarian disorders at concentrations < 100 kAU/L. Receiver-operating characteristic (ROC) curve analysis of ovarian neoplasma vs perimenopausal controls indicated a gain of specificity and sensitivity with the improved IMx assay over the RIA, but ROC performance was the same with either assay if patients with benign ovarian disorders were used as controls.
منابع مشابه
Analytical and Clinical Performance of Improved Abbott lMx CA 125 Assay: Comparison with Abbott CA 125 RIA
We compared the improved Abbott lMx cancer antigen (CA) 125 assay (cat. no. 7A89) with the Abbott CA 125 RIA. Serum specimens were from healthy perimenopausal women (n = 124) and from patients with benign gynecologic and nongynecologic diseases (n = 124), ovarian carcinoma (n = 104), or other malignancies (n = 193). The lMx assay detected as little as 0.193 kAU/L CA 125 (AU = arbitrary Abbott u...
متن کاملPerformance characteristics of seven automated CA 125 assays.
Cancer antigen 125 (CA 125) is a high-molecular-mass glycoprotein that is used as a tumor marker to monitor disease progression and response to therapy and in early detection of recurrence after treatment for ovarian cancer. The Access 2 (Beckman Coulter, Brea, CA), ADVIA Centaur (Bayer Diagnostics, Tarrytown, NY), ARCHITECT i2000 (Abbott Diagnostics, Abbott Park, IL), AxSYM (Abbott Diagnostics...
متن کاملAn enzyme immunoassay procedure for cancer antigen 125 evaluated.
The performance of a new enzyme immunoassay (EIA) procedure (Abbott Labs.) for cancer antigen 125 (CA 125) met or exceeded the manufacturer's claims for all analytical variables examined. Overall correlation with results obtained with a radioimmunoassay (RIA) were good. However, near the decision thresholds typically chosen to define a positive result for ovarian carcinoma, EIA results were 10 ...
متن کاملIMx (Abbott) Immunoassay of Insulin: A Practical Alternative to RIA Hyperinsulinemia Identification in Idiopathic Neurootology and Other Hyperinsulin Metabolic Disorders.
Hyperinsulinemia identification as defined by glucose/insulin tolerance has been established as the prime etiological factor in idiopathic neurootological disorders. Insulin assays by radioimmunoassay (RIA) and the IMx (Abbott) immunoassay yielded in 558 of 595 glucose/insulin tolerance a concurrence of 93.7%. The latter measures insulin without cross-reaction with proinsulin. The RIA methodolo...
متن کاملComparison of methods for measuring free thyroxin in nonthyroidal illness.
Patients with severe nonthyroidal illness (NTI) often have decreased serum thyroxin (T4) concentrations and sometimes have decreased free T4 (FT4) and increased tumor necrosis factor-alpha (TNF-alpha) concentrations. We evaluated four commercial methods for measuring FT4 [Abbott IMx, Amersham Amerlex MAB, Nichols, and Diagnostic Products Corporation (DPC) RIA] and one method for TNF-alpha in 41...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 41 2 شماره
صفحات -
تاریخ انتشار 1995