Methods and perceptual thresholds for short-term electrical stimulation of human retina with microelectrode arrays.
نویسندگان
چکیده
PURPOSE To report methods for performing epiretinal electrical stimulation with microfabricated electrode arrays and determining perceptual thresholds on awake human volunteers during acute surgical trials. METHODS Four hypotheses were tested: (1) epiretinal stimulation can be performed during acute experiments without obviously damaging the retina or degrading vision or the health of the eye; (2) perception can be obtained 50% of the time in blind patients with charge densities below published safety limits; (3) the minimal charge needed to induce perception would be higher in patients with more severe retinal degeneration; and (4) threshold charge would be lower at shorter stimulus durations. Five subjects with severe blindness from retinitis pigmentosa and one with normal vision (who underwent enucleation of the eye because of orbital cancer) were studied. Electrical stimulation of the retina was performed on awake volunteers by placing a single 250-microm diameter handheld needle electrode or a 10-microm thick microfabricated array of iridium oxide electrodes (400-, 100-, or 50-microm diameter) on the retina. Current sources outside the eye delivered charge to the electrodes. Assessment of damage was made by observing the clinical appearance of the eyes, comparing pre- and postoperative visual acuity, obtaining retinal histology in one case, and comparing perceptual thresholds with published safety limits. RESULTS No clinically visible damage to the eye or loss of vision occurred. Even at sites removed from stimulation, histology revealed swollen photoreceptor inner and outer segments, which were believed to be nonspecific findings. Percepts could not be reliably elicited with 50-microm diameter electrodes using safe charges in one blind patient. With the two larger electrodes, only the normal-sighted patient had thresholds at charge densities below 0.25 and 1.0 millicoulombs (mC)/cm(2) for 400- and 100-microm diameter electrodes, respectively, which is one seemingly reasonable estimate of safety derived from the product of charge per phase and charge density per phase. In blind patients, thresholds always exceeded these levels, although most were close to these limits in patient 6. The range of charge density thresholds with the 400- microm electrode in blind patients was 0.28 to 2.8 mC/cm(2). The normal-sighted patient had a threshold of 0.08 mC/cm(2) with a 400-microm electrode, roughly one quarter of the lowest threshold in the blind patients. Strength-duration curves obtained in two blind patients revealed the lowest threshold charge at the 0.25- or 1.0-ms stimulus duration. CONCLUSIONS Threshold charge densities in severely blind patients were substantially higher than that in a normal-sighted patient. Charge densities in blind patients always exceeded one seemingly reasonable estimate of safe stimulation. The potential adversity of long-term stimulation of the retina by a prosthesis has yet to be determined.
منابع مشابه
Multi-electrode arrays technology for the non-invasive recording of neural signals: a review article
The recording of electrophysiological activities of brain neurons in the last half-century has been considered as one of the effective tools for the development of neuroscience. One of the techniques for recording the activity of nerve cells is the multi-electrode arrays (MEAs). Microelectrode arrays (MEAs) are usually employed to record electrical signals from electrogenic cells like neurons o...
متن کاملPerceptual efficacy of electrical stimulation of human retina with a microelectrode array during short-term surgical trials.
PURPOSE This work is part of a feasibility assessment of a retinal prosthesis as a means to restore vision to patients with blindness caused by retinitis pigmentosa. The primary goal was to assess the concordance of the form of induced perception and the pattern of electrical stimulation of the retina, and the reproducibility of the responses. METHODS Five volunteers with severe retinitis pig...
متن کاملTHE EFFECT OF POSTERIOR CEREBRAL PULMONARY DIRECT ELECTRICAL STIMULATION (TDCS) ON IMPROVING SPATIAL, VISUAL, AND VERBAL PERCEPTUAL ABILITIES
Background & Aims: Direct electrical stimulation of the brain is a therapeutic technique that can be effective in improving visual, verbal, and spatial perception. The present study investigated the effect of direct electrical stimulation (tDCS) of the posterior parietal cortex on improving spatial, visual, and verbal perceptual abilities. Materials & Methods: In this quasi-experimental study,...
متن کاملFlexible Electrode Array for Retinal Stimulation
In this Work, ITO/PET (Indium Tin Oxide / Polyethylene Terephthalate) electrode structure which provides biocompatibility, mechanical stability and flexibility is fabricated. Flexible ITO/PET implantable electrode array for a retina has been developed. The electrode array is fabricated on a thin PET/ITO substrate and is encapsulated using, SU-8, an insulating material. PET substrate and SU-8 po...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Investigative ophthalmology & visual science
دوره 44 12 شماره
صفحات -
تاریخ انتشار 2003