A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator

نویسندگان

  • Weiliang Du
  • James N. Yang
  • Eric L. Chang
  • Dershan Luo
  • Mary Frances McAleer
  • Almon Shiu
  • Mary K. Martel
چکیده

A quality assurance (QA) procedure was developed to evaluate the congruence between the cone-beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be placed exactly at the radiation isocenter through precalibrated room lasers or light field crosshairs. The only requirement was that the BB phantom be in a stationary position near the radiation isocenter during the image acquisition process. The radiation isocenter was determined with respect to the center of the BB using a Winston-Lutz test. The CBCT image center was found to have excellent short-term positional repro-ducibility (i.e., less than 0.1 mm of wobble in each of the x (lateral), y (vertical), and z (longitudinal) directions) in 10 consecutive acquisitions. Measured over a seven-month period, the CBCT image center deviated from the radiation isocenter by 0.40 ± 0.12 mm (x), 0.43 ± 0.04 mm (y), and 0.34 ± 0.14 mm (z). The z displacement of the 3D CBCT image center was highly correlated (ρ = 0.997) with that of the 2D kV portal image center. The correlation coefficients in the x and y directions were poor (ρ = 0.66 and -0.35, respectively). Systematic discrepancies were found between the CBCT image center and the 2D MV, kV portal image centers. For the linear accelerator studied, we detected a 0.8 mm discrepancy between the CBCT image center and the MV EPID image center in the anterior-posterior direction.This discrepancy was demonstrated in a clinical case study where the patient was positioned with CBCT followed by MV portal verification. The results from the new QA procedure are useful for guiding high-precision patient positioning in stereotactic body radiation therapy.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers *

To investigate the effects of sagging shifts of three on-board kV imaging systems (OBI) on the isocenter positioning accuracy and image quality of cone-beam CT (CBCT). A cubical phantom having a metal marker in the center that can be aligned with the radiation isocenter was used to measure sagging shifts and their variation with gantry angle on three Varian linacs with kV on-board imaging syste...

متن کامل

Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space

Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...

متن کامل

Usability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter

Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...

متن کامل

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

متن کامل

The colossal circumvention of the lung lesion during lung stereotaxy

This is a case report on stereotaxic (Stereotactic Body Radiotherapy-SBRT) for lung cancer located in the left lower lobe (Segment 6, S6). There have been no reports on marked displacement of the peripheral lung cancer during radiotherapy. A pulmonary nodule was discovered on computed tomography (CT) conducted for a persistent cough in an 87-year-old male. According to diagnostic imaging, this ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2010