Toward standardization of cardiac troponin I measurements part II: assessing commutability of candidate reference materials and harmonization of cardiac troponin I assays.
نویسندگان
چکیده
BACKGROUND Cardiac tropoin I (cTnI) measurements show an approximately 20- to 40-fold difference between assays, and better standardization and harmonization are needed. Toward this goal, the AACC cTnI Standardization Committee collaborated with the National Institute of Standards and Technology (NIST) in an earlier study to select 2 candidate reference materials (cRMs). METHODS Two troponin cRMs, a troponin C-troponin I-troponin T (CIT) complex from human heart tissue and a CIT complex from recombinant technology, were supplied to NIST for assessment of composition and purity, and cTnI value assignment. These cRMs and 6 cTnI-positive human serum pools were shipped to manufacturers of 15 cTnI assays. Commutability of the materials was examined by determining the numerical relationship for the cRM preparations between each manufacturer-specified field method and each of the other 14 field methods. These relationships were then compared with the corresponding numerical relationships for the human serum pools. Harmonization of methods was accomplished by determining regression parameters relative to the analytical system yielding values closest to the median for each serum pool. These regression parameters were used to recalculate pool values to harmonize the assays. Interassay CVs before and after harmonization were determined. RESULTS Characterization of the CIT and CI cRMs showed that these materials were of specified composition. The proportion of cTnI methods that demonstrated commutability for the CIT cRM was 45%; for the CI cRM, 39% of methods demonstrated commutability. Interassay cTnI variability for the field methods ranged from 82% to 97%, median 88%. After harmonization, variability of the serum pools for the cTnI methods was decreased to between 9.0% and 23%, median 15.5%. CONCLUSIONS The proportion of methods demonstrating commutability was too low for use as a common calibrator for the cTnI field methods. However a simple strategy using serum pools can improve harmonization of field cTnI methods by more than 5-fold. The CIT cRM was selected by the AACC cTnI standardization committee, and a new lot has been classified as the cTnI certified reference material Standard Reference Material 2921 by NIST.
منابع مشابه
Characterization of a new certified reference material for human cardiac troponin I.
BACKGROUND To address the continuing need for the standardization of clinical human cardiac troponin I (cTnI) assays, NIST, with the assistance of the AACC/IFCC Cardiac Troponin I Standardization Committee, has developed a new certified reference material, Standard Reference Material (SRM) 2921: Human Cardiac Troponin Complex. METHODS The concentration of cTnI in SRM 2921 was determined throu...
متن کاملStandardisation of cardiac troponin I measurement: past and present.
The laboratory measurement of cardiac troponin (cTn) concentration is a critical tool in the diagnosis of acute myocardial infarction (MI). Current cTnI assays produce different absolute troponin numbers and use different clinical cut-off values; hence cTnI values cannot be interchanged, with consequent confusion for clinicians. A recent Australian study compared patient results for seven cTnI ...
متن کاملValues of troponin T and myoglobin predictive of non-cardiac ischemia in rats
Objective(s):Biochemical markers are important for the timely diagnosis and follow-up of ischemic events. Most of the markers have been previously studied in the context of cardiac ischemia. However, research on markers of non-cardiac events has been insufficient. Therefore, we investigated the relationship between troponin and myoglobin which are commonly used markers of cardiac ischemia, in n...
متن کاملStandardization of cardiac troponin I assays: round Robin of ten candidate reference materials.
BACKGROUND Cardiac troponin I (cTnI) results vary 100-fold among assays. As a step toward standardization, we examined the performance of 10 candidate reference materials (cRMs) in dilution studies with 13 cTnI measurement systems. METHODS Solutions of 10 cTnI cRMs, each characterized by NIST, were shipped to the manufacturers of 13 cTnI measurement systems. Manufacturers used their respectiv...
متن کاملAnalysis of the relation between coping ways with stress and Cardiac Biomarker Troponin I in coronary heart disease patients
Introduction: Coping ways with stress in coronary heart disease patients can lead to significant changes in levels of Cardiac Biomarker Troponin I, but researchers had not explored it empirically. So, the main objective of this study was to identify the relationship between coping ways with stress and Cardiac Biomarker Troponin I and also, finding coping ways that predict changes in the concent...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 52 9 شماره
صفحات -
تاریخ انتشار 2006