Stability-Indicating RP-HPLC Method for Determination of Tamsulosin HCL in Pharmaceutical Dosage Form
نویسندگان
چکیده
A selective, specific and sensitive stability-indicating high-performance liquid chromatographic method was developed and validated for the determination of Tamsulosin in in pharmaceutical dosage forms. Celecoxib was used as Internal Standard (IS). The chromatographic conditions comprised of a reversed-phase Lichrocart / Lichrosphere C18 column (250 × 4.0 mm packed with 5) with mobile phase consisting of a mixture of Acetonitrile: T.D.W. in the ratio (40: 60). Flow rate was 0.8 mL / min. Detection was carried out at 275 nm. The retention time of Tamsulosin HCl and Celecoxib were found to be 1.608 and 2.767min respectively and the linear regression analysis data for the calibration plots showed good linear relationship in the concentration range 1 - 200 g/mL. The value of correlation coefficient, slope and intercept were, 0.9995, 0.7453 and 0.4584, respectively. Tamsulosin HCl was subjected to stress conditions of degradation in aqueous solutions including acidic, alkaline, oxidation, photolysis and thermal degradation. The developed method was validated with regard to linearity, accuracy, precision, selectivity and robustness and the method was found to be precise, accurate, linear and specific. The method was employed successfully for identification and determination of Tamsulosin in pharmaceutical preparations.
منابع مشابه
Stability Indicating Rp-hplc Method Development & Validation for Simultaneous Determination of Dutasteride and Tamsulosin in Bulk as Well as in Pharmaceutical Dosage Form by Using Pda Detector
Objective: The present work was undertaken with the aim to develop and validate a rapid and consistent stability indicating RP-HPLC in which the peaks will be appear with short period of time as per ICH Guidelines. The proposed method was simple, fast, accurate and precise method for the Quantification of drug in the dosage form, bulk drug as well as for routine analysis in Quality control. Met...
متن کاملA Simple and Specific Stability- Indicating RP-HPLC Method for Routine Assay of Adefovir Dipivoxil in Bulk and Tablet Dosage Form
A simple and reliable stability-indicating RP-HPLC method was developed and validated for analysis of adefovir dipivoxil (ADV).The chromatographic separation was performed on a C18 column using a mixture of acetonitrile-citrate buffer (10 mM at pH 5.2) 36:64 (%v/v) as mobile phase, at a flow rate of 1.5 ml/min. Detection was carried out at 260 nm and a sharp peak was obtained for ADV at a reten...
متن کاملA Simple and Specific Stability- Indicating RP-HPLC Method for Routine Assay of Adefovir Dipivoxil in Bulk and Tablet Dosage Form
A simple and reliable stability-indicating RP-HPLC method was developed and validated for analysis of adefovir dipivoxil (ADV).The chromatographic separation was performed on a C18 column using a mixture of acetonitrile-citrate buffer (10 mM at pH 5.2) 36:64 (%v/v) as mobile phase, at a flow rate of 1.5 ml/min. Detection was carried out at 260 nm and a sharp peak was obtained for ADV at a reten...
متن کاملAnalytical Method Development & Validation for Simultaneous Determination of Dutasteride and Tamsulosin in Bulk as Well as in Pharmaceutical Dosage Form by Using Rp-hplc
Objective: The present work was undertaken with the aim to develop and validate a rapid and consistent RP-HPLC method in which the peaks will be appear with short period of time as per ICH Guidelines. Methods: The proposed method was simple, fast, accurate and precise method for the Quantification of drug in the dosage form, bulk drug as well as for routine analysis in Quality control. Reversed...
متن کاملA New Stability Indicating RP-HPLC Method Development and Validation for the Simultaneous Estimation of Dolutegravir and Rilpivirine in Bulk and its Dosage Forms
A New Stability Indicating RP-HPLC Method Development and Validation for the Simultaneous Estimation of Dolutegravir and Rilpivirine in Bulk and its Dosage Forms. Abstract The objective of the work is to develop and validate a new, simple, highly sensitive, stability indicating RP-HPLC method for simultaneous estimation of Dolutegravir and Rilpivirine in bulk and its dosage f...
متن کامل