Evaluation of peak inspiratory pressure, tidal volume and respiratory rate during ventilation of premature lambs using a self-inflating bag.
نویسندگان
چکیده
OBJECTIVE To evaluate the peak inspiratory pressure, tidal volume and respiratory rate achieved during manual ventilation of premature lambs, using a self-inflating bag. METHODS In this descriptive, experimental study, five pairs of physicians, selected at random among 35 neonatologists working at a neonatal intensive care unit and with experience in the resuscitation of newborn infants, ventilated five intubated premature lambs using a self-inflating bag. Pressure and flow monitor signals were passed through a transducer and digitized for recording and analysis. Tidal volume and pressure curves were obtained from the integral of flow rate, at peak, during the last 50 seconds of every fifth minute, and analyzed. RESULTS Median pressure was 39.8 (IQ(25-75%) 30.2-47.2) cmH(2)O; being below 20 in 1.1% of cases and above 40 in 49.1%. Seven out of 10 physicians produced more than six pressure peaks of over 40 cmH(2)O. Median tidal volume/kg was 17.8 (IQ(25-75%) 14.1-22.4) mL, being below 5 mL in 0.1% of cases and greater than or equal to 20 mL in 37.7%. All of the physicians propelled five or more ventilation cycles with tidal volume/kg of 20 mL or more. Respiratory rate was between 30 and 60 cycles/minute in 65.9% of cases, being below 30 in 6.8% of cases and over 60 in 27.3% of cases. CONCLUSIONS There was major variation in peak inspiratory pressure and tidal volume/kg values, which were in many cases elevated, attaining levels that habitually cause biotrauma, while respiratory rates were adequate in the majority of cases.
منابع مشابه
Effects of decreasing inspiratory flow rate during simulated basic life support ventilation of a cardiac arrest patient on lung and stomach tidal volumes.
If the airway of a cardiac arrest patient is unprotected, basic life support with low rather than high inspiratory flow rates may reduce stomach inflation. Further, if the inspiratory flow rate is fixed such as with a resuscitator performance may improve; especially when used by less experienced rescuers. The purpose of the present study was to assess the effect of limited flow ventilation on r...
متن کاملNeonatal and pediatric manual hyperinflation: influence of oxygen flow on ventilation parameters.
BACKGROUND Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags. METHODS We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactur...
متن کاملOptimizing bag-valve-mask ventilation with a new mouth-to-bag resuscitator.
When ventilating an unintubated patient with a self-inflating bag, high peak inspiratory flow rates may result in high peak airway pressure with subsequent stomach inflation; this may occur frequently when rescuers without daily experience in bag-valve-mask ventilation need to perform advanced airway management. The purpose of this study was to assess the effects of a newly developed self-infla...
متن کاملEvaluation of peak inspiratory pressure and respiratory rate during ventilation of an infant lung model with a self-inflating bag.
OBJECTIVE To evaluate the peak inspiratory pressure and ventilation rate achieved by physicians when using a neonatal self-inflating bag on a lung model. METHODS Fifteen physicians ventilated full term and preterm infant lung simulators while the outcomes were captured by a ventilation monitor. RESULTS Median peak pressures in cmH(2)O for full term and preterm lungs were 23 (interquartile r...
متن کاملAn evaluation of peak inspiratory pressure, tidal volume, and ventilatory frequency during ventilation with a neonatal self-inflating bag resuscitator.
BACKGROUND Although the self-inflating bag is widely used in the hospital setting, variability of delivered ventilatory parameters is usually high, which might result in both hypoventilation and lung injury. The aims of this study were to assess possible sources of the high variability and to evaluate the adequacy of obtained values in relation to the recommended values for neonatal resuscitat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Jornal de pediatria
دوره 82 4 شماره
صفحات -
تاریخ انتشار 2006