Long-Term Follow-Up Results of a Second-Generation Cementless Femoral Prosthesis with a Collar and Straight Distal Fixation Channels

نویسندگان

  • Chang Wook Han
  • Ick Hwan Yang
  • Hye Yeon Lee
  • Chang Dong Han
چکیده

PURPOSE We evaluated the results of more than 10 years of follow-up of total hip arthroplasty using a second-generation cementless femoral prosthesis with a collar and straight distal fixation channels. MATERIALS AND METHODS One hundred five patients (129 hips) who underwent surgery between 1991 and 1996 for primary total hip arthroplasty using cementless straight distal fluted femoral stems were followed for more than 10 years. Ninety-four hips in 80 patients were available for clinical and radiologic analysis. The mean age at the time of surgery was 47 years, and the mean duration of follow-up was 14.3 years. RESULTS The mean Harris hip scores had improved from 58 points to 88 points at the time of the 10-year follow-up. Activity-related thigh pain was reported in nine hips (10%). At the last follow-up, 93 stems (99%) were biologically stable and one stem (1%) was revised because of loosening. No hip had distal diaphyseal osteolysis. Proximal femoral stress-shielding was reported in 86 hips (91%). We found no significant relationship between collar-calcar contact and thigh pain, stem fixation status, or stress-shielding. The cumulative survival of the femoral stem was 99% (95% confidence interval, 98-100%) after 10 years. CONCLUSION The long-term results of total hip arthroplasty using a second-generation cementless femoral prosthesis with a collar and straight distal fixation channels were satisfactory; however, the high rate of proximal stress-shielding and the minimal effect of the collar indicate the need for some changes in the stem design.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cementless Hip Arthroplasty in Southern Iran, Midterm Outcome and Comparison of Two Designs

Background: Cementless hip prosthesis was designed to provide biologic fixation, without the use of cement. The second generation components have shown more reliable bone ingrowths and survival rates. We are reporting a midterm result of two designs of cementless prosthesis in a unique culture with different social habits and expectations.Methods: 52 primary cementless total hip arthroplasty in...

متن کامل

Long-term Results of Osteoarticular Allograft Reconstruction in Children with Distal Femoral Bone Tumors

Background: There is no consensus regarding the best method of reconstruction in pediatric population following thewide resection of malignant bone tumors. More exploration of the complications of osteoarticular reconstruction leads toless existing controversy of this type of reconstruction, which is the main point of this article.Methods: Long-term outcomes and complications of osteoarticular ...

متن کامل

Revision Fixation of Distal Humerus Fracture Nonunions in Older Age Patients with Poor Bone Quality or Bone Loss – Is This Viable as a Long-term Treatment Option?

Background: The purpose of this retrospective study was to analyze the long-term results of revision ORIF, jointcontracture release, and autogenous bone-grafting in the treatment of distal humerus frac-ture nonunions in older agedpatients with poor bone quality or bone loss who would have been candidates for total elbow arthroplasty.Methods: Seven patients (average age at inde...

متن کامل

Cementless Femoral Fixation in Total Hip Arthroplasty

In 1995, the NIH (National Institutes of Health) Consensus Development Panel on Total Hip Replacement supported the use of hybrid fixation with a cemented stem and a noncemented cup because of excellent long-term results. However, today, 60% to 90% of the approximately 200,000 total hip arthroplasties performed yearly in the United States involve both a cementless cup and a cementless stem. The...

متن کامل

Primary total hip arthroplasty without the use of bone cement: a 10-year follow-up of 157 hips.

BACKGROUND The cementless fixation technique in total hip arthroplasty (THA) was developed to solve clinical problems such as aseptic loosening and osteolysis which were thought to be associated with the use of bone cement. This retrospective study reports our mid-term results with cementless THA. METHODS A series of 173 consecutive, unselected cementless THA procedures using the Omnifit pros...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 53  شماره 

صفحات  -

تاریخ انتشار 2012