Preconcentration of Environmental Waters by Agar for XRF Analysis
نویسندگان
چکیده
We have developed a convenient and effective XRF analysis procedure for trace amount of K, Ca, V, Mn, Fe, Ni, Cu, Zn, Cd, and Pb in environmental waters by using a preconcentration using the natural polymer (agar). The thin agar film was prepared by drying a homogeneous agar gel after mixing the aqueous sample solution with the agar powder. XRF analysis of the preconcentrated thin agar films containing trace metals showed a good repeatability, because agar films were homogeneous enough. SB (signal to noise) ratios of the XRF intensity of the analytes were improved drastically. The linear calibration curves of K, Ca, V, Mn, Fe, Ni, Cu, Zn, Cd, and Pb showed a good linearity within the calibration ranges. The lower limits of detection (LLD) were 1.4 μg/mL for K, 0.26 μg/mL for Ca, 0.088 μg/mL for V, 0.029 μg/mL for Mn, 0.11 μg/mL for Fe, 0.016 μg/mL for Ni, 0.030 μg/mL for Cu, 0.017 μg/mL for Zn, 0.20 μg/mL for Cd, and 0.066 μg/mL for Pb, respectively. The proposed preconcentration method was applied to several environmental water samples. 96 Copyright ©JCPDS-International Centre for Diffraction Data 2009 ISSN 1097-0002
منابع مشابه
Effect of oxidation of activated carbon on its enrichment efficiency of metal ions: comparison with oxidized and non-oxidized multi-walled carbon nanotubes.
The effect of oxidation of activated carbon (AC) with various oxidizing agents (nitric acid, hydrogen peroxide, ammonium persulfate) on preconcentration of metal ions (Cr3+, Mn2+, Pb2+, Cu2+, Cd2+ and Zn2+) from environmental waters prior to their flame atomic absorption spectroscopic analysis was investigated. The highest recoveries and adsorption capacities towards metal ions were achieved wh...
متن کاملPartially pyrolyzed olive pomace sorbent of high permeability for preconcentration of metals from environmental waters.
The aim of this work is to develop a preconcentration procedure of Cd(2+), Zn(2+) and Cu(2+) in environmental waters using olive pomace (OP) prior to their determination by flame atomic absorption spectrometry (FAAS). Raw OP as preconcentrating sorbent was found to have low permeability towards the passed water samples and thus long time was needed. Even reducing the vacuum pressure caused cart...
متن کاملEffect of Chemical Treatment of Multi-Walled Carbon Nanotubes with Various Oxidizing Agents on its Preconcentration Performance of Some Metals
Multi-walled carbon nanotube was oxidized with various oxidizing agents (nitric acid, hydrogen peroxide and ammonium persulphate) to produce multi-walled carbon nanotubes of various surface properties. The oxidized sorbents were used for the preconcentration of some metals (Mn, Pb, Cu, Cd, Zn) from environmental waters prior to their determination by flame atomic absorption spectroscopy. The be...
متن کاملUltrasound Assisted Emulsification Microextraction as a Simple Preconcentration Method for Determination of Atrazine in Environmental Samples
We represent an environmentally friendly sample of pre-treatment method, ultrasound assisted emulsification microextraction (USAEME) followed by gas chromatography nitrogen phosphorous detector (GC-NPD), to specify atrazine residues in environmental samples. Some parameters affecting the extraction efficiency such as the type and volume of the extraction solvent, emulsification time and additio...
متن کاملPreconcentration and Determination of Chromium (III) from Sea Water Samples Using Ion Imprinted Activated Carbon
A simple ion imprinted amino-functionalized sorbent was synthesized by coupling activated carbon with iminodiacetic acid, a functional compound for metal chelating, through cyanoric chloride spacer. The resulting sorbent has been characterized using Fourier transform infrared spectroscopy (FT-IR), elemental analysis, thermogravimetric analysis (TGA) and evaluated for the preconcentration and de...
متن کامل