Preliminary Retrospective Analysis of Daily Tomotherapy Output Constancy Checks Using Statistical Process Control
نویسندگان
چکیده
The purpose of this study was to retrospectively evaluate the results from a Helical TomoTherapy Hi-Art treatment system relating to quality controls based on daily static and dynamic output checks using statistical process control methods. Individual value X-charts, exponentially weighted moving average charts, and process capability and acceptability indices were used to monitor the treatment system performance. Daily output values measured from January 2014 to January 2015 were considered. The results obtained showed that, although the process was in control, there was an out-of-control situation in the principal maintenance intervention for the treatment system. In particular, process capability indices showed a decreasing percentage of points in control which was, however, acceptable according to AAPM TG148 guidelines. Our findings underline the importance of restricting the acceptable range of daily output checks and suggest a future line of investigation for a detailed process control of daily output checks for the Helical TomoTherapy Hi-Art treatment system.
منابع مشابه
Retrospective analysis of linear accelerator output constancy checks using process control techniques
Shewhart control charts have previously been suggested as a process control tool for use in routine linear accelerator (linac) output verifications. However, a comprehensive approach to process control has not been investigated for linac output verifications. The purpose of this work is to investigate a comprehensive process control approach to linac output constancy quality assurance (QA). The...
متن کاملOnce Daily Accelerated Partial Breast Irradiation: Preliminary Results with Helical Tomotherapy®.
BACKGROUND Accelerated partial breast irradiation (APBI) is becoming an option for patients with low-risk breast cancer. The current practice is 38.5 Gy in 10 fractions b.i.d. over 5 days. This fractionation has a higher bioequivalent dose compared to the standard schedule. We report on preliminary results of once-daily APBI in patients treated with TomoTherapy®. PATIENTS AND METHODS Patients...
متن کاملEvaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter‐free (FFF) photon beams
Machine Performance Check (MPC) is an automated and integrated image-based tool for verification of beam and geometric performance of the TrueBeam linac. The aims of the study were to evaluate the MPC beam performance tests against current daily quality assurance (QA) methods, to compare MPC performance against more accurate monthly QA tests and to test the sensitivity of MPC to changes in beam...
متن کاملElectron beam quality control using an amorphous silicon EPID.
An amorphous silicon EPID has been investigated to determine whether it is capable of quality control constancy measurements for linear accelerator electron beams. The EPID grayscale response was found to be extremely linear with dose over a wide dose range and, more specifically, for exposures of 95-100 MU. Small discrepancies of up to 0.8% in linearity were found at 6 MeV (8-15 MeV showed bet...
متن کاملSystematic and comprehensive analysis of the dose-response characteristics of a morning quality check of a linear accelerator and an important application of accelerator performance prediction
Background: This paper aimed to analyze the output constancy of a medical linear accelerator using PTW QUICKCHECKwebline and studied the sensitivity of the PTW QUICKCHECKwebline. Materials and Methods: The paper statistically analyzed the output doses of 6 MV and 10 MV of photons and 6 MeV, 9 MeV, and 12 MeV of electrons from a medical linear accelerator measured before the daily treatment to ...
متن کامل