QDR 4500A DXA overestimates fat-free mass compared with criterion methods.
نویسندگان
چکیده
This study evaluated the accuracy with which the dual-energy X-ray absorptiometer (Hologic QDR 4500A) measured fat-free mass (FFM), fat mass (FM), and hydration of FFM. In a study of 58 men and women (ages 70-79 yr), the QDR 4500A was found to provide a systematically higher estimate of FFM and lower estimate of FM than a four-component model of body composition. A correction factor from this study was developed and applied to two other samples (n = 13 and 37). We found mean corrected levels of FFM and FM to be equivalent to that obtained by the four-component model or total body water. In addition, the hydration of the corrected FFM was closer to the established hydration level in adult samples and that obtained from the four-component model. These findings suggest that the current calibration of the fan-beam system of the Hologic QDR 4500A provides an overestimate of FFM and underestimate of FM compared with reference methods.
منابع مشابه
Measurement of fat mass using DEXA: a validation study in elderly adults.
The accuracy of total body fat mass and leg fat mass measurements by fan-beam dual-energy X-ray absorptiometry (DEXA) was assessed in 60 healthy elderly subjects (aged 70-79 yr). Total fat and leg fat mass at four leg regions (total leg, thigh, midthigh, and calf) were measured with the QDR 4500A (Hologic, Waltham, MA). The four-compartment model and multislice computed tomography scans were se...
متن کاملCross-calibration of pencil-beam (DPX-NT) and fan-beam (QDR-4500C) dual-energy X-ray absorptiometry for sarcopenia
Sarcopenia, defined as the loss of muscle mass accompanied by weakness, is an important factor leading to frailty and is a growing concern in the aging Japanese society. Muscle mass can be calculated by dual-energy X-ray absorptiometry (DXA), but results differ between devices produced by different manufactures. Thus, cross-calibration is needed to compare body composition results in multicente...
متن کاملComparison of total body potassium with other techniques for measuring lean body mass in men and women with AIDS wasting.
BACKGROUND Lean body mass is an important predictor of survival and functional status in patients with AIDS wasting. The bias between different techniques for assessing body composition in AIDS wasting is not known. DESIGN We compared total body potassium (TBK) with fat-free mass (FFM) determined by dual-energy X-ray absorptiometry (DXA), bioelectrical impedance analysis (BIA), and skinfold-t...
متن کاملDual-energy X-ray absorptiometry for the measurement of gross body composition in rats.
Dual-energy X-ray absorptiometry (DXA) is a novel, non-invasive technique for the measurement of gross body composition in small animals. In the present study the absolute accuracy of the Hologic QDR-1000W scanner was assessed by comparison with direct analysis in twelve rats with a range of body fat and bone mineral content (BMC) values. Fat masses measured by DXA and petroleum-ether extractio...
متن کاملNormative Data for Bone Mass in Healthy Term Infants from Birth to 1 Year of Age
For over 2 decades, dual-energy X-ray absorptiometry (DXA) has been the gold standard for estimating bone mineral density (BMD) and facture risk in adults. More recently DXA has been used to evaluate BMD in pediatrics. However, BMD is usually assessed against reference data for which none currently exists in infancy. A prospective study was conducted to assess bone mass of term infants (37 to 4...