Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer.
نویسندگان
چکیده
Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as well as the automated count, included metamyelocytes, myelocytes, and promyelocytes. Manual differentials were performed on each specimen by 2 experienced medical technologists using the National Committee for Clinical Laboratory Standards (NCCLS) document H20-A. In addition to evaluating the IG parameter, St. Anthony's laboratory plans to report the automated IG count from the XE-2100, with or=5% and an absolute IG number of >or=0.5 x 103 will require differentiation by a manual differential count. An IG percentage <5% and an absolute IG number of <0.5 x 103 will not require differentiation by manual slide review. In this comparison of the IG count on the XE-2100 to the manual count, the correlation coefficient was 0.83. The XE-2100 IG results showed significant correlation with the results from the manual immature granulocyte count. We concluded that the XE-2100 IG demonstrated comparable performance to the manual IG count.
منابع مشابه
Evaluation of the Diagnostic Performance of theSysmex XT-2000i Automated Hematology Analyzerin the Detection of Immature Granulocyte
Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microsco...
متن کاملPreliminary Data on the Feasibility of Bone Marrow Screening on the Sysmex XE-2100 Automated Hematology Analyzer
In the present study we investigated the potential of the Sysmex XE-2100 automated hematology analyzer as a screening device for bone marrow (BM) aspirates. We evaluated BM cellularity, white blood cell (WBC) differential count, immature granulocyte (IG) count, nucleated red blood cell (NRBC) count, granulocyte/erythroid (G/E) and myeloid/erythroid (M/E) ratio, and potential pathologyrelated sc...
متن کاملComparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples.
Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in ...
متن کاملEvaluation of immature granulocyte counts by the XE-IG master: upgraded software for the XE-2100 automated hematology analyzer.
We evaluated an automated immature granulocyte (IG) count in the peripheral blood with the XE-IG Master (Sysmex Corporation). The XE-IG Master is the upgraded software package for the XE-2100 automated hematology analyzer. Reproducibility tests demonstrated a mean coefficient of variation of 7.02% for the IG percentage (IG%) and 6.93% for the absolute IG count. Correlations of the IG counts wer...
متن کاملReference Values for Immature Granulocytes in Healthy Blood Donors Generated on the Sysmex XE-2100 Automated Hematology Analyer
An increase in immature granulocytes (IG) in peripheral blood is of great importance for the diagnosis of hematological malignancies. Furthermore, it may indicate systemic inflammation. IGs are usually quantified by visual microscopy; however, recent studies have evaluated automated IG count on the DIFF channel and IG detection on the IMI channel of the Sysmex XE-2100 automated hematology analy...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Laboratory hematology : official publication of the International Society for Laboratory Hematology
دوره 12 1 شماره
صفحات -
تاریخ انتشار 2006