Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies

نویسندگان

  • Roman Hovorka
  • Kavita Kumareswaran
  • Julie Harris
  • Janet M Allen
  • Daniela Elleri
  • Dongyuan Xing
  • Craig Kollman
  • Marianna Nodale
  • Helen R Murphy
  • David B Dunger
  • Stephanie A Amiel
  • Simon R Heller
  • Malgorzata E Wilinska
  • Mark L Evans
چکیده

OBJECTIVE To compare the safety and efficacy of overnight closed loop delivery of insulin (artificial pancreas) with conventional insulin pump therapy in adults with type 1 diabetes. DESIGN Two sequential, open label, randomised controlled crossover, single centre studies. SETTING Clinical research facility. PARTICIPANTS 24 adults (10 men, 14 women) with type 1 diabetes, aged 18-65, who had used insulin pump therapy for at least three months: 12 were tested after consuming a medium sized meal and the other 12 after consuming a larger meal accompanied by alcohol. INTERVENTION During overnight closed loop delivery, sensor measurements of glucose were fed into a computer algorithm, which advised on insulin pump infusion rates at 15 minute intervals. During control nights, conventional insulin pump settings were applied. One study compared closed loop delivery of insulin with conventional pump therapy after a medium sized evening meal (60 g of carbohydrates) at 1900, depicting the scenario of "eating in." The other study was carried out after a later large evening meal (100 g of carbohydrates) at 2030, accompanied by white wine (0.75 g/kg ethanol) and depicted the scenario of "eating out." MAIN OUTCOME MEASURES The primary outcome was the time plasma glucose levels were in target (3.91-8.0 mmol/L) during closed loop delivery and a comparable control period. Secondary outcomes included pooled data analysis and time plasma glucose levels were below target (≤ 3.9 mmol/L). RESULTS For the eating in scenario, overnight closed loop delivery of insulin increased the time plasma glucose levels were in target by a median 15% (interquartile range 3-35%), P = 0.002. For the eating out scenario, closed loop delivery increased the time plasma glucose levels were in target by a median 28% (2-39%), P = 0.01. Analysis of pooled data showed that the overall time plasma glucose was in target increased by a median 22% (3-37%) with closed loop delivery (P < 0.001). Closed loop delivery reduced overnight time spent hypoglycaemic (plasma glucose ≤ 3.9 mmol/L) by a median 3% (0-20%), P=0.04, and eliminated plasma glucose concentrations below 3.0 mmol/L after midnight. CONCLUSION These two small crossover trials suggest that closed loop delivery of insulin may improve overnight control of glucose levels and reduce the risk of nocturnal hypoglycaemia in adults with type 1 diabetes. Trial registration ClinicalTrials.gov NCT00910767 and NCT00944619.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Unsupervised home use of an overnight closed-loop system over 3–4 weeks: a pooled analysis of randomized controlled studies in adults and adolescents with type 1 diabetes

AIMS To compare overnight closed-loop and sensor-augmented pump therapy in patients with type 1 diabetes by combining data collected during free-living unsupervised randomized crossover home studies. METHODS A total of 40 participants with type 1 diabetes, of whom 24 were adults [mean ± standard deviation (s.d.) age 43 ± 12 years and glycated haemoglobin (HbA1c) 8.0 ± 0.9%] and 16 were adoles...

متن کامل

Home use of closed-loop insulin delivery for overnight glucose control in adults with type 1 diabetes: a 4-week, multicentre, randomised crossover study.

BACKGROUND Closed-loop insulin delivery is a promising option to improve glycaemic control and reduce the risk of hypoglycaemia. We aimed to assess whether overnight home use of automated closed-loop insulin delivery would improve glucose control. METHODS We did this open-label, multicentre, randomised controlled, crossover study between Dec 1, 2012, and Dec 23, 2014, recruiting patients from...

متن کامل

Closed-loop artificial pancreas using subcutaneous glucose sensing and insulin delivery and a model predictive control algorithm: the Virginia experience.

BACKGROUND Recent progress in the development of clinically accurate continuous glucose monitors (CGMs), automated continuous insulin infusion pumps, and control algorithms for calculating insulin doses from CGM data have enabled the development of prototypes of subcutaneous closed-loop systems for controlling blood glucose (BG) levels in type 1 diabetes. The use of a new personalized model pre...

متن کامل

Closed-loop artificial pancreas using subcutaneous glucose sensing and insulin delivery and a model predictive control algorithm: preliminary studies in Padova and Montpellier.

New effort has been made to develop closed-loop glucose control, using subcutaneous (SC) glucose sensing and continuous subcutaneous insulin infusion (CSII) from a pump, and a control algorithm. An approach based on a model predictive control (MPC) algorithm has been utilized during closed-loop control in type 1 diabetes patients. Here we describe the preliminary clinical experience with this a...

متن کامل

Day-and-night glycaemic control with closed-loop insulin delivery versus conventional insulin pump therapy in free-living adults with well controlled type 1 diabetes: an open-label, randomised, crossover study

BACKGROUND Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve glucose control while alleviating the risk of hypoglycaemia in adults with HbA1c below 7·5% (58 mmol/mol). METHODS In this open-label, randomised, crossover study, we recruited adults ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 342  شماره 

صفحات  -

تاریخ انتشار 2011