Accuracy of in-line venous saturation and hematocrit monitors in pediatric perfusion.

نویسندگان

  • S M Yaskulka
  • J Burnside
  • D Bennett
  • V Olshove
  • J Langwell
چکیده

Cardiopulmonary bypass (CPB) in neonates and infants routinely employs lower blood flow rates (BFR), temperatures, and hematocrits (Hct) than those typically observed in adult CPB. The purpose of this study was to evaluate the accuracy of three devices available for continuous in-line measurement of venous oxygen saturation (SvO2) and Hct during pediatric CPB. Venous blood samples were obtained over a range of BFR, temperatures, and Hct and analyzed on a Corning 2500 Co-Oximeter and HematoStat C-70 centrifuge. These values were then compared to those measured by the Bentley OxySat SM-0200, the Gish StatSat, and the 3M CDI 100 in-line monitors. Seventy samples were obtained and analyzed using linear regression, paired t-test and residual analysis to establish the reliability and accuracy of each device. The results demonstrate the CDI to be more statistically accurate (p > 0.05) than the Gish and Bentley devices for measuring SvO2 in nearly all circumstances, though all correlated well with the control (r > 0.70). When comparing spun Hct to the CDI and Gish values, the CDI in-line monitor demonstrated a greater reliability to predict actual patient Hct (r > 0.90) than the Gish StatSat (r > 0.60). Residual analysis revealed that even though the Gish StatSat had higher calculated p values (p > 0.05) than the CDI 100 for interpreting Hct, it was shown to display more inconsistent and sporadic values over the ranges of BFR and temperature studied. It is concluded that the CDI 100 proved to be more accurate, reliable, and consistent than the Gish StatSat and the Bentley Oxysat devices in determining SvO2 and Hct over all evaluated parameters in this study.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical comparisons of continuous venous oxygen saturation and hematocrit monitors in pediatric surgery.

Continuous venous oxygen saturation and hematocrit values are important parameters in assessing patient status while on cardiopulmonary bypass. Two devices used to measure continuous venous oxygen saturation while on cardiopulmonary bypass were compared to a control. The Bentley Oxysat meter and the Medtronic MX2 Oxygen Saturation and Hematocrit System were compared to the ABL500 blood gas mon...

متن کامل

In vitro evaluation of continuous mixed venous oxygen saturation and hematocrit monitors.

Four venous hemoglobin oxygen saturation monitors, two of which provide hemoglobin monitoring, were evaluated and compared to a performance standard (control) during cardiopulmonary bypass (CPB). The CDI 100 Hematocrit/Saturation Monitor, CDI 400, Cobe Hematocrit/Saturation Monitor, and the Baxter Bentley OxySat values were compared to the Radiometer Copenhagen OSM3 Hemoximeter and the Damon In...

متن کامل

Continuous venous oximetry: a comparative study between the CDI 100 and the Bentley OxySat II.

Two in-line oxygen saturation monitors, the CDI 100 and OxySat II, were evaluated in the clinical setting. Eighty-seven venous blood samples were drawn during 20 elective cardiopulmonary bypass procedures. Monitor readings were compared to OSM III co-oximeter values. The results revealed that saturation (%) determination was biased, -3.16 +/- 2.21 SD for the CDI 100 and -0.34 +/- 2.17 SD for th...

متن کامل

Validity of Peripheral Capillary Oxygen Saturation and Normal Mixed Venous Oxygen Tension in Measurement of O2 Content

Background The change in venous oxygen saturation occurs earlier, and even its reduction is faster than arterial oxygen saturation. The aim of this study was to validate SvO2 and PvO2 for O2 content measurement in children hospitalized w...

متن کامل

Effects of Hypothermic Cardiopulmonary Bypass on Internal Jugular Bulb Venous Oxygen Saturation, Cerebral Oxygen Saturation, and Bispectral Index in Pediatric Patients Undergoing Cardiac Surgery: A Prospective Study

The objective of this study was to evaluate the effect of hypothermic cardiopulmonary bypass (CPB) on cerebral oxygen saturation (rSO2), internal jugular bulb venous oxygen saturation (SjvO2), mixed venous oxygen saturation (SvO2), and bispectral index (BIS) used to monitor cerebral oxygen balance in pediatric patients.Sixty American Society of Anesthesiologists Class II-III patients aged 1 to ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 27 3  شماره 

صفحات  -

تاریخ انتشار 1995