Evaluation of the Doppler technique for fat emboli detection in an experimental flow model.

نویسندگان

  • Victoria Wikstrand
  • Nadja Linder
  • Karl Gunnar Engström
چکیده

Pericardial suction blood (PSB) is known to be contaminated with fat droplets, which may cause embolic brain damage during cardiopulmonary bypass (CPB). This study aimed to investigate the possibility to detect fat emboli by a Doppler technique. An in vitro flow model was designed, with a main pump, a filter, a reservoir, and an injector. A Hatteland Doppler probe was attached to the circulation loop to monitor particle counts and their size distribution. Suspended soya oil or heat-extracted human wound fat was analyzed in the model. The concentrations of these fat emboli were calibrated to simulate clinical conditions with either a continuous return of PSB to the systemic circulation or when PSB was collected for rapid infusion at CPB weaning. For validation purpose, air and solid emboli were also analyzed. Digital image analysis was performed to characterize the nature of the tested emboli. With soya suspension, there was an apparent dose response between Doppler counts and the nominal fat concentration. This pattern was seen for computed Doppler output (p = .037) but not for Doppler raw counts (p = .434). No correlation was seen when human fat suspensions were tested. Conversely, the image analysis showed an obvious relationship between microscopy particle count and the nominal fat concentration (p < .001). However, the scatter plot between image analysis counting and Doppler recordings showed a random distribution (p = .873). It was evident that the Doppler heavily underestimated the true number of injected fat emboli. When the image analysis data were subdivided into diameter intervals, it was discovered that the few large-size droplets accounted for a majority of total fat volume compared with the numerous small-size particles (< 10 microm). Our findings strongly suggest that the echogenecity of fat droplets is insufficient for detection by means of the tested Doppler method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

شناسایی میکرو آمبولی مغزی در سیگنال داپلر از روی ویژگی های غیر خطی

Abstract: An embolus is a blood clot, a fat globule or gas bubbles that may be freely circulating in bloodstream can stop the blood flow and lead to ischemia. In real time assessment of blood flow by Trans Cranial Doppler (TCD) method, travelling solid or gaseous micro emboli in the bloodstream by passing across the assessment area, causes a short time signal with high intensity. While TCD reco...

متن کامل

I-46: Obstetrical Doppler

Accurate assessment of gestational age, fetal growth, and the detection of fetal and placental abnormalities are major benefits of sonography. Color Doppler can be used to assist in the identification of vascular architecture, detection of vascular pathology and visualization of blood flow changes associated with physiologic processes and disease states. The clinical applications of obstetrical...

متن کامل

Detection of arterial emboli using Doppler ultrasound in rabbits.

The purpose of this study was to develop an animal model that could be used to test the ability of Doppler ultrasound to detect arterial emboli composed of materials that are often involved in cerebral emboli. Emboli introduced into the rabbit aorta via the left renal artery consisted of clotted whole blood, platelets, atheromatous material, fat, or air. The ultrasound examination was carried o...

متن کامل

Detection of Middle Cerebral Emboli During Coronary Artery Bypass Surgery Using Transcranial Doppler Sonography

Detection of Middle Cerebral Emboli During Coronary Artery Bypass Surgery Using Transcranial Doppler Sonography To the Editor: Spencer et al recently described the use of transcranial Doppler sonography to record the passage of emboli through the middle cerebral artery (MCA) during carotid endarterectomy. They recorded transient signals at the time of releasing common carotid cross clamps and i...

متن کامل

Transcranial Doppler Detection of Circulating Cerebral Emboli

Background: The identification of gaseous emboli using Doppler ultrasound was described as early as the 1960s. Recently it has been demonstrated that this method can also detect solid emboli such as thrombi and platelet aggregates. This may make this technology useful in a large number of patients with, or at risk of, embolic stroke. Summary of Review: Emboli appear as short-duration, high-inte...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 40 3  شماره 

صفحات  -

تاریخ انتشار 2008