CT venography after knee replacement surgery: comparison of dual-energy CT-based monochromatic imaging and single-energy metal artifact reduction techniques on a 320-row CT scanner

نویسندگان

  • Masafumi Kidoh
  • Daisuke Utsunomiya
  • Seitaro Oda
  • Takeshi Nakaura
  • Yoshinori Funama
  • Hideaki Yuki
  • Kenichiro Hirata
  • Masahiro Hatemura
  • Tomohiro Namimoto
  • Yasuyuki Yamashita
چکیده

BACKGROUND An optimal metal artifact reduction (MAR) technique is needed for a reliable and accurate image-based diagnosis. PURPOSE Using a 320-row scanner, we compared the dual-energy computed tomography (CT)-based monochromatic and the single-energy metal artifact reduction (SEMAR) techniques for CT venography (CTV) to identify the better imaging method for diagnosing deep vein thrombosis (DVT) in patients who had undergone knee replacement surgery. MATERIAL AND METHODS Twenty-three consecutive patients with suspected DVT after unilateral knee replacement surgery underwent dual-energy CT (135/80 kVp). Monochromatic images of 35-135 keV were generated; the monochromatic image with the best signal-to-noise ratio (SNR) of the popliteal vein near the metal prosthesis were selected. The projection data of 80 kVp were reconstructed using MAR algorithm. The mean SNR ON MAR and the best SNR ON monochromatic images were compared. Two radiologists evaluated visualization of the metal artifacts on a four-point scale where 1 = extensive artifacts, 2 = strong artifacts, 3 = mild artifacts, and 4 = minimal artifacts. RESULTS The mean SNR was significantly higher on the MAR than the monochromatic images (12.8 ± 4.7 versus 7.7 ± 5.1, P < 0.01) and the visual scores were significantly higher for MAR than monochromatic images (2.6 ± 0.8 versus 1.3 ± 0.4, P < 0.01). CONCLUSION For CTV after knee replacement surgery, the MAR technique is superior to the monochromatic imaging technique.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Deep Vein Thrombosis with Multidetector Row CT after Orthopedic Arthroplasty: a Prospective Study for Comparison with Doppler Sonography

OBJECTIVE This prospective study evaluated the ability of indirect 16-row multidetector CT venography, in comparison with Doppler sonography, to detect deep vein thrombosis after total hip or knee replacement. MATERIALS AND METHODS Sixty-two patients had undergone orthopedic replacement surgery on a total of 30 hip joints and 54 knee joints. The CT venography (scan delay time: 180 seconds; sl...

متن کامل

AbstractID: 8312 Title: The use of dual-energy CT images for Monte Carlo treatment planning: material extraction and metal artifact reduction

ID: 8312 Title: The use of dual-energy CT images for Monte Carlo treatment planning: material extraction and metal artifact reduction Purpose: To investigate the feasibility of dual-energy CT-based material extraction (DECT) for Monte Carlo (MC) dose calculations, to evaluate metal streaking artifact reduction using DECT and to propose a novel scanning technique for motion artifacts reduction. ...

متن کامل

Evaluation of Metal Artifact Reduction software in Computed Tomography

Introduction: The image quality of computed tomography (CT) can be seriously lowered by metal implants of patients. These implants are known to exert a significant impact on diagnostic accuracy due to artifacts. The current study aimed to assess the usefulness of Metal Artifact Reduction (MAR) software in the reduction of metal artifacts, in comparison to iterative rec...

متن کامل

A novel dual energy CT-based attenuation correction method in PET/CT systems: A phantom study

  In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-en...

متن کامل

Comparing 511 keV Attenuation Maps Obtained from Different Energy Mapping Methods for CT Based Attenuation Correction of PET Data

Introduction:  The  advent  of  dual-modality  PET/CT  scanners  has  revolutionized  clinical  oncology  by  improving lesion localization and facilitating treatment planning for radiotherapy. In addition, the use of  CT images for CT-based attenuation correction (CTAC) decreases the overall scanning time and creates  a noise-free  attenuation  map  (6map).  CTAC  methods  include  scaling,  s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 6  شماره 

صفحات  -

تاریخ انتشار 2017