Atypical fungal obstruction of the left ventricular assist device outflow cannula
نویسندگان
چکیده
Background Even if better clinical outcomes have been achieved with the implantation of newer generation continuous-flow left ventricular assist devices (LVADs), infection complications are still a major risk for these patients in longterm follow-up. We present a case of a 56-year-old male with dilated cardiomyopathy who was urgently implanted with left ventricular assist device HeartMate II. During the entire postoperative period, we observed repeated life-threatening septic complications. Clinical signs of infection disappeared after prolonged treatment with broad-spectrum antibiotics and patient was discharged to outpatient monitoring. During the entire duration of circulatory support, no LVAD suction events were detected, pump power consumption remained in the normal range and the patient was listed for heart transplantation. Patient was two month rehospitalized due to worsening of his status. Hemoglobinuria, an increase of inflammatory markers, and alteration in hepatic and renal function were detected. TEE imaging showed no obstructive formation in inflow or outflow cannulas; only velocities were decreased. The LV was severely dilated and the aortic valve was opening at each contraction. Patient was initially considered for systemic thrombolysis, but this urgent status was finally solved with explantation of the LVAD and subsequent heart transplantation. Massive obstruction with thrombuslike formation was found in the outflow cannula. Histopathologic examination and microbiologic culturing of the mass showed extensive fungal growth (Aspergillus species). During the postoperative period, patient was aggressively treated with antifungal therapy. He was discharged 8 weeks after the transplantation. Conclusion We believe that clinical decision-making in patients with signs of ongoing sepsis and end-organ dysfunction together with any signs of LVAD malfunction should be very straightforward in terms of device exchange or explantation.
منابع مشابه
Case report: atypical fungal obstruction of the left ventricular assist device outflow cannula
We describe a very rare case of outflow cannula obstruction with fungal infectious thrombus formation. Discussion includes the etiology, diagnosis, and management of fungal infection complications related with long-term mechanical circulatory support. Left ventricular assist devices (LVADs) are increasingly used as bridge to transplant and permanent long-term therapy in the population with end-...
متن کاملVADoscopy: a novel intraoperative technique to evaluate HeartMate II left ventricular assist device inflow obstruction and thrombosis.
With HeartMate II (HMII) implants increasing so has the frequency of device exchange. However, identifying inflow cannula obstruction, pump thrombosis, or outflow obstruction as the mechanism of pump dysfunction can be difficult. Echocardiography, CT angiogram, and cardiac catheterization are not definitive in determining the location of pump failure. Therefore, intraoperative examination is of...
متن کاملLeft ventricular assist device outflow cannula obstruction treated with percutaneous endovascular stenting.
T hrombosis within the circuit of a left ventricular assist device (LVAD) is a major complication imparting significant morbidity and mortality. Recent management algorithms emphasize anticoagulation or surgery for LVAD thrombosis, depending on clinical presentation. 1 We report a case of thrombotic LVAD outflow cannula obstruction treated successfully with percutaneous endovascular stenting. A...
متن کاملLeft ventricular non-compaction cardiomyopathy and left ventricular assist device: a word of caution
BACKGROUND In patients with left ventricular non-compaction (LVNC), implantation of a left ventricular assist device (LVAD) may be performed as a bridge to transplantation. In this respect, the particular characteristics of the left ventricular myocardium may represent a challenge. CASE PRESENTATION We report a patient with LVNC who required urgent heart transplantation for inflow cannula obs...
متن کاملIn vivo evaluation of the Nimbus axial flow ventricular assist system. Criteria and methods.
Continuing in vivo trials are being conducted at the University of Pittsburgh using the Nimbus axial flow blood pump (AxiPump). To date, 14 sheep experiments have been performed to address several issues related to short-term support. Six acute experiments (< 6 hr) have been performed to assess hemodynamics related to speed regulation and to determine anatomic placement of the pump and cannulae...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 8 شماره
صفحات -